ActiveCare - Standard
User Manual
MAN-ActiveCare REV03
ActiveCare Standard USER MANUAL
Page 30 of 33
Bed Transport
To transport an unoccupied bed, unlock the casters and use the footboard or headboard to move
the bed, do not move the bed using the rails, handheld pendant or assist bars as handles. When
bed is moved into intended space, lock casters.
NOTE: The bed is not intended to provide a means for transport of the occupant.
CAUTION: Never operate bed while casters are unlocked.
Maintenance
Cleaning
The frame can be cleaned/disinfected with a disinfecting wipe. Disinfectant should always be
applied to a cloth; never spray directly on frame or controls. After applying cleaning agent, allow
the
surface to dry prior to placing unit back in service. Clean frame as needed.
Maintenance and Service
NOTICE: RECOMMENDED TO PERFORM PREVENTIVE MAINTENANCE EVERY 12
MONTHS FOR BEST PERFORMANCE.
1. Inspect the frame for weld cracks or tears.
2. Check that all bolts, pins and fasteners are still in working order.
3. Perform functional test on the frame.
4. Inspect the casters for wear and tear.
Preventative Maintenance Inspections
Problem Resolution
Inspect bed frame for cracks, tears or bending,
particularly at welds and joints
.
Remove bed from service. Contact Med-Mizer®
customer service.
Inspect electrical enclosures for cracks or other
damage
.
Remove bed from service. Contact Med-Mizer®
customer service.
Check that screws, bolts, pins and other fasteners are
in place, secured and properly tightened
.
Remove bed from service. Tighten loose bolts and
screws. Contact Med-Mizer® customer service to order
missing or damaged fasteners.
Inspect actuators for excessive wear and
misalignment
.
Remove bed from service. Contact Med-Mizer®
customer service.
Inspect power cord for damage including the plug
and strain relief
.
Remove bed from service. Contact Med-Mizer®
customer service to order replacement power cord.