8
silverlight - manual of use and maintenance
04.0 Installation
04.1 Safety requirements at the time of installation
�
DANGER:
The wiring system of the premises where the apparatus is installed and used must comply
with the applicable standards and the relevant electrical safety requirements.
�
DANGER:
Do not install the apparatus in places where there is a risk of explosion. The apparatus
may not be used in areas where there are inflammable atmospheres (anesthetic mixtures, oxygen, etc.).
�
DANGER:
Install the apparatus in a place where it will be protected from blows and from accidental
sprays of water or other liquids.
�
DANGER:
Do not install the apparatus above or in the vicinity of sources of heat. Make sure there is
sufficient air circulating around the apparatus.
�
DANGER:
Do not short circuit the electric contacts of the recharging unit with metal objects
(Fig.5 - Ref.B) and do not touch them with your hands while the apparatus is switched on.
�
WARNING:
The apparatus is transportable, however it must be handled with care when moved.
�
WARNING:
Do not expose the apparatus to direct sunlight or sources of UV light.
04.2 Description of the controls and
signaling lamps
Description of the controls (Fig. 4):
Ref. A
Green power LED.
Function:
This indicates that the recharging unit is
powered up.
Ref. B
Battery LED.
Function
Green:
Indicates that the battery of the
Silverlight is being charged.
Red:
Indicates that the battery of the
Silverlight has failed.
Ref. C
Test LED.
Function
Green:
Indicates that the light intensity is
suitable for effective therapy.
Red:
Indicates that the light intensity is
insufficient.
Ref. D
Push-button for activating and cutting off
the emission of light.
Function:
This starts or stops a polymerization cycle.
A
D
C
B
Fig. 4