32
8.1.4
Specifications of the tests for the Immunity of the Accessible
Parts of the Casing to the Wireless RF Communications Device
PIEZOSURGERY® GP is designed to operate in an electromagnetic environment in which
radiated RF disturbances are under control. The customer or the user of PIEZOSURGERY® GP
can help prevent electromagnetic interference by ensuring a minimum distance between the
mobile and portable RF (transmitters) communication devices and PIEZOSURGERY® GP, as
recommended, in relation to the maximum output power of radiocommunications equipment.
Test
Freq.
(MHz)
(MHz)
Service
Max
power
(W)
Distance
(m)
Immunity
test level
(V/m)
385
380 - 390
TETRA 400
Pulse
18 Hz
1.8
0.3
27
450
430 - 470
GMRS 460
FRS 460
FM
± 5 kHz
deviation
1 kHz sine
2
0.3
28
710
704 - 787
LTE band 13,
17
Pulse
217 Hz
0.2
0.3
9
745
780
810
800 - 960
GSM 800/900
TETRA 800
iDEN 820
CDMA 850
Band LTE 5
Pulse
18 Hz
2
0.3
28
870
930
1720
1700 - 1990
GSM 1800
CDMA 1900
GSM 1900
DECT
LTE Band 1, 3,
4, 25
UMTS
Pulse
217 Hz
2
0.3
28
1845
1970
2450
2400 - 2750
Bluetooth
WLAN
802.11 b/g/n
RFID 2450
Band LTE 7
Pulse
217 Hz
2
0.3
28
5420
5100 - 5800
WLAN
802.11 a/n
Pulse
217 Hz
0.2
0.3
9
5500
5785
a)
For some services, only the uplink frequencies are
included.
b)
The carrier shall be modulated using a 50 % duty cycle
square wave signal.
c)
As an alternative to FM modulation, 50 % pulse
modulation at 18 Hz may be used because while it
does not represent actual modulation, it would be
worst case.