15
USE
FUNCTIONS (Ref. P front endpaper)
Depending on the type of clinical
application, it is possible to choose
one of the 6 options available from the
“function” list:
•
ENDO
: dedicated to endodontic
surgery, to Schneider’s membrane
detachment and conventional
endodontic treatments;
•
PERIO
: dedicated to periodontal
surgery and conventional scaling
and periodontal treatments;
•
SPECIAL
: dedicated only to the
inserts for osteotomies with
thickness of 0.35 mm and for
prosthetic applications;
•
CANCELLOUS
: dedicated to the
cutting and removal of poorly
mineralized bone;
•
CORTICAL
: dedicated to the cutting
and removal of highly mineralized
bone;
•
IMPLANT
: dedicated to bone
perforation in the technique of the
implant site preparation.
FILLING
AND
FLUSHING
THE
IRRIGATION CIRCUIT (Ref. Q front
endpaper)
The device is equipped with the “PUMP/
FLUSH” key which, depending on the
mode of use, allows to perform the PUMP
function or the FLUSH function.
The PUMP function can be used at the
beginning of the treatment, to flush the
entire irrigation line up to the insert, so
that the surgery can be started with the
necessary irrigation (see Chapter 5.3 on
The FLUSH function allows to run a
flushing cycle of the irrigation circuit
of the handpiece(s) used during the
treatment.
pump/flush
function
irrigation
6
5
4
3
2
1
0
implant
cortical
cancellous
special
perio
endo