20
CAUTION
: Do not twist or rotate the
foot pedal cable connector when inserting
or removing it. Twisting can damage the
connector.
5.3 Instructions for Use
After having connected all the accessories as described in
as follows:
Open the air inlet on the spike, before
proceeding with the operation.
Open the irrigation tube clamps/roller, if
closed;
NOTE:
Before using the device, check
that the irrigation tube clamps/roller are
open. After surgery, close the clamps/
roller before disconnecting the irrigation
kit from the physiological irrigation bag.
1
To fill the irrigation circuit, use the PUMP
function.
To activate the FLUSH function, press and
hold the button PUMP/FLUSH on the
touch keyboard till the irrigation liquid
outflows from the handpiece: all the
other buttons present on the console’s
keyboard are disabled and the LED of the
level 6 of the “irrigation” function starts
flashing.
NOTE:
The PUMP/FLUSH cycle can be
interrupted at any time by releasing the
PUMP/FLUSH key on the console’s touch
keyboard.
2
pump/flush
function
irrigation
6
5
4
3
2
1
0
implant
cortical
cancellous
special
perio
endo
Screw the chosen insert onto the
PIEZOSURGERY® GP handpiece, till the
end of the threading;
3