![Mecmesin Multitest 0.5d Service And Repair Manual Download Page 9](http://html.mh-extra.com/html/mecmesin/multitest-0-5d/multitest-0-5d_service-and-repair-manual_1759033009.webp)
Multitest 0.5d, 1d & 2.5d
Service and Repair Manual
Number 022-005
Issue Number: 08
Page 9 of 58
8. Transformer removal and Replacement Parvalux
7.
Remove the base plate as per 7, Instructions 2 to 5
8.
Locate the transformer as
identified in the photograph
9.
Remove the
grey/blue/brown/mauve
transformer connections to the
mains inlet filter
10. Remove the yellow/orange
transformer connections to the
PCB
NOTE - record the connections before
removing
11. Using a 5mm ball ended allen key remove the 2-off M6x12 socket cap screws
retaining the transformer mounting plate to the base casing
12. Lift out the transformer. Unscrew nylon nut, Remove transformer from plate.
NOTE
: The transformer
must
be insulated from its mounting plate. An insulating
gasket is supplied with the transformer for this purpose
–
make sure it is fitted
Refitting of the Transformer is the reverse of the above procedure