Assembly and Installation of –
i
System Test Stands
Mecmesin
17
Vortex-
i
0.3 N m
1.5 N m
3 N m
6 N m
10 N m
Measurement range
N.m
kgf.cm
lbf.in
0-0.3
0-3
0-2.7
0-1.5
0-15
0-13
0-3
0-30
0-26
0-6
0-60
0-52
0-10
0-100
0-90
Load Measurement
Load accuracy
±0.5% of full scale
Load resolution
1:6500
Load units
mN.m, N.cm, N.m, kgf.cm, gf.cm, oz.fin, lbf.ft, lbf.in
Speed
Speed range
0.1-20 rev/min (clockwise or anticlockwise)
Speed accuracy
±1% of indicated speed
Speed resolution
±0.1 rev/min
Displacement
Maximum displacement
2440 revs
Displacement accuracy
0.2º per 36,000º
Displayed resolution
0.001 revs (±0.2º)
Dimensions
Max travel of adjustable
transducer carriage
182 mm (7.2’)
Maximum headroom
505 mm (19.9’) [448 mm (17.6’)]*
Width between columns
208 mm (11.02’)
Weight
19.5 kg (43 lbs)
Capacity of lower mounting table
10-190 mm (0.39-7.5’)
Capacity of upper mounting table
10-78 mm (0.39-3.07’)
Miscellaneous
Max power requirements
100 W
Voltage
230 V AC 50 Hz or 110 V AC 60 Hz
Loadcell calibration temperature
20±2°C
* With upper and lower mounting tables fitted