approximately five minutes has elapsed since the last key press, the last load change greater than 2% or the last
recurrence of RS-232 communication.
Auto-off
If the gauge is being used for a long test the user may wish to disable the auto-off function. The auto-off function
can be disabled when powering up the gauge by holding down the
until the display shows ‘No Ao’.
Overload
Do not overload
Do not overload
the loadcell, as this can cause irreparable damage.
Forces greater than 120% of full-scale will show an ‘-OL-‘ symbol and the corresponding tension or compression
symbols on the display. If the display shows ‘OL’ (without dashes) in either of the two max modes, the loadcell is
not currently in overloaded state but has been overloaded at least once since the last
ZERO
ZERO
operation.
An instrument showing an overload condition cannot be relied upon to provide accurate, repeatable
An instrument showing an overload condition cannot be relied upon to provide accurate, repeatable
measurement - consult your supplier.
measurement - consult your supplier.
Overload counters
If you suspect that the gauge has been subjected to overloading this can be verified by pressing the
UNITS
UNITS
key
while turning on the gauge. This will first display the number of overloads in tension, press
UNITS
UNITS
key again to see
the number of overloads in compression. Press
UNITS
UNITS
key once more to allow the gauge to resume powering up.
Removing recorded overloads can only be undertaken by Mecmesin or an approved distributor.
Should the instrument have sustained a catastrophic overload the symbol ‘OL’ will be permanently displayed and
the instrument must be returned to Mecmesin or an approved Mecmesin distributor for repair.
Display messages
‘-OL-‘
Overload - More than 120% of full scale load is
currently being applied to the transducer
‘OL’
Overload - The peak force reading was applied to
the loadcell in excess of 120% of full-scale
‘t-ERR’
Tare Error - The zero function was performed while
the transducer was in an overloaded state
‘No Ao’
No Auto off - Automatic power-off is disabled
‘C-dEF’
Calibration default - Invalid calibration data; please
notify Mecmesin or an authorised representative for
calibration
RS232 Commands Table: Configuration
It is possible to remotely read/configure the settings of your CFG+ by sending the following RS232 command
characters:
12