3.6
3.6.1
3.6.2
3.6.3
3.6.4
Character in
Character in
ASCII
ASCII
Decimal
Decimal Hexadecimal
HexadecimalFunction
Function
M
77
0x4D
Current mode
U
85
0x55
Current units
C
67
0x43
Loadcell capacity in newtons
?
63
0x3F
Transmit the current reading
@
64
0x40
Configuration Status Request
*
42
0x2A
Toggle Continuous Transmit
Mode
r
114
0x72
Normal Screen (Non max)
t
116
0x74
Max Tension Screen
u
117
0x75
Max Compression Screen
a
97
0x61
Change Units to mN
b
98
0x62
Change Units to N
c
99
0x63
Change Units to kN
d
100
0x64
Change Units to gf
e
101
0x65
Change Units to kgf
f
102
0x66
Change Units to ozf
g
103
0x67
Change Units to lbf
CTRL a
1
0x01
Simulates the
TXD
TXD
key
CTRL b
2
0x02
Simulates the
UNITS
UNITS
key
CTRL c
3
0x03
Simulates the
MAX
MAX
key
CTRL e
5
0x05
Simulates the
ZERO
ZERO
key (to
ZERO a load reading, not
turn the gauge off)
CTRL o
15
0x0F
Turns the gauge off
RS232 Command Responses: Information
It is possible to remotely interrogate your CFG+ by sending the following RS232 commands. This will inform you
which settings are currently configured.
Command: M
Response
Response
CFG+
CFG+ Display Mode
Display Mode
Normal
Normal Mode
MaxC
Max Compression
MaxT
Max Tension
Command: U
Response
Response
mN
kN
N
gf
kgf
ozf
lbf
Command: C
The loadcell size in N
Note:
Note:
‘xxxx’ will be transmitted if the loadcell is not calibrated, or has a serious fault. Contact Mecmesin or your
supplier.
Command: @