CAPTEST
Operator Manual
Mecmesin
11
Configuring
CAPTEST
Settings
Settings for the
CAPTEST
are available from the
configuration menu, and are not intended for the regular
user to access. Please note the following:
The settings you make are not visible from the display
(e.g. there is no indication of units)
There is no confirmation of the settings you choose: be
methodical.
There is no ‘exit’ from the configuration settings other than
by pressing
ENTER
and accepting a setting.
Settings are selected and made only one at a time.
Settings are retained when the unit is switched off.
The settings menu
Settings are found and saved one at a time. To enter the
settings, hold
ZERO
while pressing
ON
.
Press the
ON
key until the display shows
SEt
.
Press
ENTER
: the display will show
U1
(the first settings
item)
o
to set
U1
(use lbf.in as the units of measurement), press
ENTER
o
or to find another setting in the list, press
ON
until it
displays, and then press
ENTER
to set it.
Pressing
ON
returns you from the settings menu back to
continuous mode measurement.