Note: If you do not want to print
CAPTEST
data each time it
is sent, but only to have statistical analysis data at the end of
a test batch, disable printing using the [Printer] key.
CAPTEST
data will still be stored.
Switching off
Always switch the
CAPTEST
off first, and then the DP-1 VR.
This is important, because
CAPTEST
is equipped with an
automatic interface detection.
Detach the interface connection only if both the
CAPTEST
and DP-1VR are switched off.
Setting tolerance limits
Clear stored data: press the DP-1 VR [CL] key and clear
the
CAPTEST
memory:
ON
>
ON
>
ZERO
.
Press [TOL. LIMIT] on the DP-1 VR to begin setting values.
Apply torque to the
CAPTEST
; to the upper or lower limit
required (sequence does not matter). Transmit the value to
the DP-1 VR:
ON
>
ZERO
.
Repeat the last step to the second limit torque.
Press the [TOL. LIMIT] key of the DP-1 VR. This will
complete the setting and print the tolerance limits:
LSL=lower limit; USL=upper limit, TOL=tolerance range.
For each measurement transmitted to the DP-1 VR, red
and green indicator lights will show if the result is below
the lower limit, above the upper limit, or within limits.
When you have completed a batch of measurements, make
sure the printer is on and press the DP-1 VR [STAT] key to
print the data values and histogram.