Mechan PLA-600 User Manual Download Page 8

DOC. MEUM-131205700. Version 01

Section 3

System Sketch
3.1

Connection Graph:

1.

Host Device

2.

Power Supply Cable

3.

Footswitch Plate

4.

Footswitch for ABLATION

5.

Footswitch for PLACOAG

6.

Treatment Handle

7/8.

Plasma Probe

9.

Socket for Footswitch

10.

Socket for Treatment Handle

3.2

Working Principles

3.2.1

Principle:

This unit can transmit energy to the electrode part at the far end of treatment probe. Electricity flows
between working electrode and circuit electrode, and only generate local energy field. So this design can
generate controllable energy output and minimal collateral tissue damage. In reverse, mono polar system
only has one working electrode at the far end. Electricity comes from the working electrode, and flows
through human body to reach the electrode plate attached to patient, thus making the energy through
human body and surrounding tissues get much larger.

3.2.2

How it Works:

To transmit the energy to the electric conduction liquid (e.g. normal saline), which is close to or on the
surface of target tissues. The electric conduction liquid forms a thin layer between working electrode and
circuit electrode. The electric conduction liquid, given with enough energy, can transform to a steam layer
(i.e. plasma) containing charged particles with energy. When the high energy charged particles contact
target tissues, it will realize tissue disintegration by molecular dissociation. Compared with the
conventional electrosurgical system, by this kind of working model, the temperature generated at the target
tissues is much lower, thus bringing less collateral damage to the surrounding tissues. While the unit

1

2

3

4

5

6

7

8

9

10

Summary of Contents for PLA-600

Page 1: ...logy Co Ltd No 11 District 9 Huayin Industrial Harbor No 618 West Kelin Road Chengdu Straits Science Tech Industrial Development Park Wenjiang District Chengdu China Customer Service 400 666 0366 Emai...

Page 2: ...Electronic Technology Co Ltd with all rights reserved No part of this document may be used reproduced transmitted processed or recorded by any means or form electronic mechanical photographic or othe...

Page 3: ...CH 3 7 3 1 CONNECTION GRAPH 3 3 2 WORKING PRINCIPLE 3 4 3 2 1 Principle 3 3 2 2 How It Works 3 4 3 2 3 Drawing of Principle 4 3 3 SYSTEM STRUCTURES AND CONFIGURATION INTRODUCTION 5 7 3 3 1 Introductio...

Page 4: ...ortation 10 5 6 2 Storage 10 5 7 REGULATIONS ON SCRAPED EQUIPMENT 10 SECTION 6 SYSTEM SPECIFICATIONS 11 6 1 TECHNICAL SPECIFCIATION 11 6 1 1 Treatment Handle 11 6 1 2 Host Device 11 6 1 3 Power Output...

Page 5: ...and turning on the unit To correct it please do as follows A Change the place of unit B Put the unit farther away from the influenced devices C Connect the unit to different electrical outlet or elect...

Page 6: ...Attention Provides important matters requiring attention at first Section 1 Shows the purpose and organization of this manual Section 2 Describes the general configuration clinical indication and con...

Page 7: ...re use the handle should be disinfected And there are bipolar and multi polar probes supplied 2 2 Clinical Indication and Contraindication 2 2 1 Indication Urology Surgeries BPH etc 2 2 2 Contraindica...

Page 8: ...flows through human body to reach the electrode plate attached to patient thus making the energy through human body and surrounding tissues get much larger 3 2 2 How it Works To transmit the energy t...

Page 9: ...nit Output Interface Chip Drive Display Sound Instruction Picture 1 The picture 1 shows the main structures of this device Signals are formed in the vibration unit and then output is realized in isola...

Page 10: ...igns refers to the Mark of Flow Control Unit refers to the Mark of Protection Class ours is CF type refers to Footswitch Connection refers to Volume Adjustment at the Host Machine refers to Power Supp...

Page 11: ...sion 01 3 3 2 Control Keys and Marking of the Host Device Shown with Graph 1 Manufacturer Logo 2 Four Pads 3 Socket for Footswitch 4 Socket for Treatment Handle 2 3 16 15 14 13 4 6 5 7 8 9 10 11 12 19...

Page 12: ...output is closed operator can press the keys for change 13 Display of time After the device is started it will show the work time 14 Power Supply Display Light Green After power on the light is green...

Page 13: ...to the socket at the front of host device and rotate it tightly 4 Plug the treatment handle into the corresponding socket of the device Before use make sure the treatment handle is cleaned and disinf...

Page 14: ...il reaching the highest of 10 Gear After 10 the system will go back to 1 Gear 2 To reduce the gear position power output value press the key to reduce the voltage level of the corresponding mode Each...

Page 15: ...the marking of Upwards Handle with Care Weather Proof 5 6 2 Storage This device should be stored in a well ventilated room without corrosive gas The height above ground and the distance from wall sho...

Page 16: ...est Weight 10KG Height 19CM Width 39CM Length 45CM 6 1 5 Footswitch Cable length 4 5m Cleanness way anhydrous ethanol 6 2 System Classifications and Security Statement 6 2 1 System Classification As p...

Page 17: ...curity Statement This unit meets the requirements of Term 19 1 19 2 and 19 3 of GB9706 1 2007 and GB9706 4 2009 Leakage electricity Grounding 200 A Leakage electricity Patients 20 A The unit can be us...

Page 18: ...st Compliance Electromagnetic environment guidance 4 RF emissions CISPR 11 Group 1 The Model PLA 600 uses RF energy only for its internal function Therefore its RF emissions are very low and are not l...

Page 19: ...ospitalenvironment Surge IEC 61000 4 5 1 kV differential mode 2 kV 1 kV differential mode 2 kV common mode common mode Mains power quality should be that of a typical commercial or hospitalenvironment...

Page 20: ...es m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicin...

Page 21: ...odel PLA 600 as recommended below according tothe maximum output power of the communications equipment Rated maximum output of transmitter W Separation distance according to frequency of transmitter m...

Reviews: