DOC. MEUM-131205700. Version 01
Guidance and manufacturer's declaration – electromagnetic immunity –
for all EQUIPMENT and SYSTEMS
SYSTEMSGuidance and manufacturer´s declaration – electromagnetic immunity
The Model PLA-600 are intended for use in the electromagnetic environment specified below.
Thecustomer or the user of the Model PLA-600 should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment
-guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact ± 8
kV air
± 6 kV contact ± 8
kV air
Floors should be wood, concrete
orceramic tile. If floors are covered with
synthetic material, the relativehumidity
should be at least 30 %.
Electrostatic
transient / burst
IEC 61000-4-4
± 2 kV for power
supply lines ± 1
kV for
input/output lines
± 2 kV for power
supply lines
± 1 kV for
input/output lines
Mains power quality should be that of a
typical commercial or
hospitalenvironment.
Surge IEC
61000-4-5
± 1 kV differential
mode ± 2 kV
± 1 kV differential
mode± 2 kV
common mode common mode Mains
power quality should be that of a typical
commercial or hospitalenvironment.
Voltage dips, short
interruptions
andvoltage
variations on power
supply input lines
IEC 61000-4-11
< 5 % UT
(95 % dip in UT )
for 0.5 cycle
40 % UT
(60 % dip in UT )
for 5 cycles
70 % UT
(30 % dip in UT )
for 25 cycles
< 5 % UT
(>95 % dip in UT )
for 5 sec
< 5 % UT
(95 % dip in UT )
for 0.5 cycle
40 % UT
(60 % dip in UT )
for 5 cycles
70 % UT
(30 % dip in
UT ) for 25 cycles
< 5 % UT
(>95 % dip in UT )
for 5 sec
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of themodels
PLA-600 product namerequires
continued operationduring power mains
interruptions, it is recommended that the
models PLA-600 be powered from an
uninterruptible power supply or abattery.
Power frequency
(50/60 Hz)
magnetic field IEC
61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
NOTE: UT is the a. c. mains voltage prior to application of the test level.