DOC. MEUM-131205700. Version 01
Section 5
Instructions of Use
5.1
Requirement on Operators:
Operators should be experienced in electrosurgical surgeries, should get to know the latest progress
ofUrology surgeries.
5.2
General System Operations
5.2.1
Footswitch
:
To activate the host device functions. It has two functions as below:
1.
To activate ABLATION functions: Press the yellow switch for ABLATION to activate the
ABLATION mode.
2.
To activate PLACOAG functions: Press the blue switch for PLACOAG to activate the PLACOAG
mode.
5.2.2
Gear Position (Power Output Value) Adjustment
1.
To increase the gear position (power output value): press the key to increase the voltage level of the
corresponding mode. Each press will make the output voltage increase by one level until reaching
the highest of 10 Gear. After 10, the system will go back to 1 Gear.
2.
To reduce the gear position (power output value): press the key to reduce the voltage level of the
corresponding mode. Each press will make the output voltage reduce by one level until reaching the
lowest level of 1 Gear. After 1, the system will go back to 10 Gear.
5.3
Voltage Output
The voltage output value of ABLATION and PLACOAG is LED displayed in single numbers at the front
control panel. The highest voltage output is as follows:
1.
Highest Load Peak Value Voltage for Different Modes (Vrms)
Modes
ABLATION (Load: 250 Ohm)
PLACOAG
(Load:
250
Ohm)
Output Peak Value of the 10
th
Gear
295
158
2.
Highest Open Circuit Peak Value Voltage for Different Models (VP)
Modes
ABLATION
PLACOAG
Output Peak Value of the 10
th
Gear
680
380
5.4
System Preparations:
Refer to the installation and checkout process. And to choose probe type suitable for the actual surgery and
Output Voltage
Output Voltage