PCI-DAS6030 & PCI-DAS6032 User's Guide
Specifications
36
Table 4. Relative accuracy
– all values are (±)
Range
Relative Accuracy (µV)
Single Point
Averaged (Note 2)
±10 V
723.3
72.3
±5 V
361.6
36.2
±2 V
144.7
14.5
±1 V
72.3
7.2
±500 mV
42.2
4.2
±200 mV
26.5
2.7
±100 mV
24.1
2.4
0 to 10 V
482.2
48.2
0 to 5 V
241.1
24.1
0 to 2 V
96.4
9.6
0 to 1 V
48.2
4.8
0 to 500 mV
33.1
3.3
0 to 200 mV
24.1
2.4
0 to 100 mV
22.9
2.3
Note 2:
Averaged measurements assume averaging of 100 single-channel readings.
Relative accuracy is defined as the measured deviation from a straight line drawn between measured
endpoints of the transfer function. ADC resolution, noise and front-end non-linearity are included in
this measurement.
Table 5. Differential non-linearity specifications
All ranges
±0.5 LSB typ
±1.0 LSB max
Settling time
Settling time is defined as the time required for a channel to settle to within a specified accuracy in response to
a full-scale (FS) step. Two channels are scanned at the specified rate. A –FS DC signal is presented to
channel 1; a +FS DC signal is presented to channel 0.
Table 6. Settling time specifications
Condition
Range
±0.00076%
(±0.5 LSB)
±0.0015%
(±1 LSB)
±0.0061%(±4
LSB)
±0.012%
(±8 LSB)
Same range to same
range
±10 V
40 µS max
20 µS max
10 µS max
5 µS typ
±5 V
40 µS max
20 µS max
10 µS max
5 µS typ
±2 V
40 µS max
20 µS max
10 µS max
5 µS typ
±1 V
40 µS max
20 µS max
10 µS max
5 µS typ
±500 mV
40 µS max
20 µS max
10 µS max
5 µS typ
±200 mV
40 µS max
20 µS max
10 µS max
5 µS typ
±100 mV
40 µS max
20 µS max
10 µS max
5 µS typ
0 to 10 V
40 µS max
20 µS max
10 µS max
5 µS typ
0 to 5 V
40 µS max
20 µS max
10 µS max
5 µS typ
0 to 2 V
40 µS max
20 µS max
10 µS max
5 µS typ
0 to 1 V
40 µS max
20 µS max
10 µS max
5 µS typ
0 to 500 mV
40 µS max
20 µS max
10 µS max
5 µS typ
0 to 200 mV
40 µS max
20 µS max
10 µS max
5 µS typ
0 to 100 mV
40 µS max
20 µS max
10 µS max
5 µS typ
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com