Instruction Manual
DPG-Series Inclinometer
MEAS Deutschland GmbH
www.meas-spec.com
phone: +49-(0)231-9740-0
Hauert 13, D-44227 Dortmund, Germany
12 of 15
fax: +49-(0)231-9740-20
July 2015
6.2.2
Command Reference SETUP level
Setup level will be used for permanent settings, stored.
Commands in
setup level
Input
Output
Comment
This level will be terminated (reset)
automatically after about 10 minutes of
user inactivity.
Stop continuous
output of inclination
values
“
c
”
“c”
Stops the continuous output of angle
results or raw values, required for input
of commands, terminated by “C”,
mode permanent,
query with “ R”, see chapter 6.2.1.
Start continuous
output of values
“
C
”
Start the continuous output of angle
results or raw values,
mode permanent, valid after reset.
Set zero
“
N
”
“N”
Define 0° position. Offset is stored and
used even after reset.
Setback zero
“
n
”
“n”
Resets the offset displacement
Set lowpass filter
“
M
”,
f
“M”,
f
Input filter settings with count
f = “0”…”7”, other characters were
answered with “E”, see item 1) below.
Set output rate
“
O
”,
o
“O”,
o
Set output rate with count o = “1”...”6”,
other characters were answered with
“E”, see item 2) below.
Show internal
settings
“
*
”
“O123 M123”, CR, LF
“OffX=+00.000
OffY=+00.000”, CR, LF
“Erase Cycl SegF:
+00255”, CR, LF
“O123”: internal setting of output rate in
10ms, “M123”: internal filter setting;
both values 255 => default settings
“Off..”: stored Offset values
“Erase Cycl Seg…”: active Segment
and number of erasure procedures for
this segment
Show currently
temperature [°C]
“
T
“
“T= 00.00”, CR,LF
Readout the currently temperature in
°C. Usable at > software version 2.0.2.
Without sign, positive temperature.
With sign `- `, negative temperature.
Software reset
“
q
”
“q”
Reset the sensor
Erase both Info
segments,
(switch to default
settings)
“
E
"
“Seg A deleted” , CR, LF
“Seg B deleted” , CR, LF
Erases the info segments of flash
storage banks, InfoA and InfoB. This
set all changed user values to default
settings
Forbidden command
“
#
”
“#”
Reserved for factory communication