ANI Monitor V1 - Continouous analgesia monitoring system
MD/PRD/IN16.ANIV1 V.10 – 29 SEP 2021
16
The principle of this patch in two parts is to be placed on each side of the heart (thus on each
side of the chest) to get a cardiac vector (the axis of the average cardiac vector according to
the reference anatomical position is: forward, down and left). In this case we can apply the
dual sensor on a hairless area of the upper chest and the single sensor on the left side of the
abdomen (see next page the example with figure 4).
WARNINGS:
Use again an array already used on previous patient could lead risk of cross-contamination.
If the patient develops a cutaneous rash or others unusual symptoms, remove the array. It is
important to adapt a particular care to patients suffering from dermatological problems.
Never put electrodes on injuries.
NOTE:
The ANI Monitor V1 has been designed to work with specific disposable electrodes.
It is inadvisable using another kind of electrode.
The electrode packaging is ten per bag. The purchase of new electrodes is available through
a box with ten bags inside thus in total one hundred arrays.
Figure 4