14
Control Test
We recommend performing Control Tests to check the
performance of the system.
Control Tests should be performed:
• Before using the System for the first time,
• For practice to ensure testing technique is good,
• When opening a new vial of test strips,
• If results seem unusually high or low, based on
patient's condition,
• If a vial has been left opened or exposed to extreme
heat or cold, or humidity,
• Whenever a check on the performance of the system is
needed,
• If meter damage is suspected (meter was dropped,
crushed, wet, etc.)
Note:
It is important to perform Control Tests with more
than one level of control solution. Three levels of
McKesson TRUE METRIX® Control Solution are
available for Control Tests. Call 1-800-803-6025
for assistance in obtaining different levels of control
solution.
Caution!
Ranges printed on test strip vial label are for
Control Test results only and are not to be used
for management of patient’s blood glucose.
Do not drink control solution.