37
4. Let meter air dry thoroughly before
using to test.
5. Verify that the system is working
properly by performing an
Automatic Self-Test (see
Getting
Started
).
Stop using the meter if:
• Meter display appears cloudy or any display segments
are missing,
• Markings on meter, including back meter label, are
coming off or are missing,
• Buttons are hard to push on the meter or do not work,
• Unable to insert test strip into Test Port,
• Automatic Self-Test gives an error message.
For general questions about McKesson TRUE METRIX®
PRO products, call 1-800-777-4908. For additional
assistance, call 1-800-803-6025.
Note:
Other disinfectants have not been tested. The effect of
other disinfectants used interchangeably has not been
tested with the meter. Use of disinfectants other than
Super Sani Cloth Wipes may damage meter.
Note:
Super Sani Cloth Wipes have been tested on the
meter for a total of 10,950 cleaning and disinfecting
cycles, which is equal to cleaning and disinfecting the
meter 10 times per day for a 3 year period. The use
life of the meter is 3 years.