McKesson LUMEON 3069 Operation Manual Download Page 6

4

Infrared Tympanic Electronic Thermometer

Ear:

  In ear (EAR) mode, the display will indicate the absolute temperature without adjustment.

Oral:  

In oral (ORL) mode, the tympanic temperature is adjusted to display an oral temperature equivalent.  Oral Mode =    

 

 

Ear Mode + 0.60°C.

Axillary: 

 In axillary (AX) mode, the tympanic temperature is adjusted to display an axillary temperature equivalent.   

 

Axillary Mode = Ear Mode + 0.04°C.

** In neonates the axillary equivalence is recommended because it is the current standard in neonates.

Core:

  In core (CORE) mode, the tympanic temperature is adjusted to display the core temperature equivalent.     

 

Core Mode = Ear Mode + 1.04°C.

Rectal:

  In rectal (REC) mode, the tympanic temperature is adjusted to display the rectal temperature equivalent.     

 

Rectal Mode = Ear Mode + 1.16°C.

Probe Covers 

The Infrared Tympanic Electronic Thermometer utilizes a single use disposable probe cover during temperature 

measurement. The probe cover was designed especially for use with the Infrared Tympanic Electronic 

Thermometer. The use of probe covers produced by any company except PSS World Medical, Inc. will result in 

erroneous temperature measurement. In order to aid in infection prevention and control, always install a new 

probe cover prior to taking a temperature. The probe covers are contained in a cassette that is located in the 

thermometer base. To load a probe cover on the thermometer, firmly insert the probe tip into the probe cover. 

When installed on the thermometer, the probe cover membrane should be smooth with no holes, tears, or 

wrinkles. After the patient temperature has been taken, eject the probe cover by pressing the eject button and then 

dispose of properly.

Temperature Measurement

1.  Remove the Infrared Tympanic Electronic Thermometer from the base.

2.  Inspect the probe lens. If any debris is present, clean the probe tip per the 

directions in Section VII, Cleaning.  If the probe tip is clean, proceed to step 3.

3.  Install a probe cover by firmly inserting the probe tip into a probe cover.  After 

the probe cover is installed, the thermometer will perform a system reset. 

The thermometer will then display dashes, the equivalence mode, and the 

thermometer icon.

4.  Inspect the probe cover to make sure it is fully seated (no space between cover 

and tip base) and no holes, tears, or wrinkles are present in the plastic film.

º F C º

º

F
C º

Summary of Contents for LUMEON 3069

Page 1: ...n Scale Dimensions Revisions Rev Description Date K PMS 110Yellow Meghan Little 3069_Masterpage Tympanic Thermometers 1 1 Proportion As Provided Final Final Draft 09 03 09 PMS 110Yellow K INFRARED TYM...

Page 2: ...5 Temperature Display Toggle 5 Sleep Mode 5 Pulse Timer Mode 6 Thermometer Display Icons and Alarms 6 Biotech Mode 7 Section VI Preventative Maintenance 9 Section VII Cleaning 9 Section VIII Battery...

Page 3: ...Features Very fast temperature acquisition that meets CEN and ASTM standards see Section XI Peak Select System Temperature measurement range of 33 0 C to 42 0 C 91 4 F to 107 6 F Once temperature has...

Page 4: ...mperatures during cold weather conditions allow the patient to equilibrate to room temperature before use Excessive eardrum scarring may cause lowered temperature readings Under normal conditions ear...

Page 5: ...Thermometer probe covers could jeopardize the accuracy of the Infrared Tympanic Electronic Thermometer Check your probe cover carton for the LumeonTM logo to be certain that your thermometer will deli...

Page 6: ...PSS World Medical Inc will result in erroneous temperature measurement In order to aid in infection prevention and control always install a new probe cover prior to taking a temperature The probe cov...

Page 7: ...age Temperature Recall After a temperature has been acquired the thermometer will enter off mode after approximately 10 seconds Recall the temperature by pressing and releasing the scan button or by p...

Page 8: ...talled or the batteries have been changed the thermometer will perform a system reset The thermometer will perform an internal test to verify that the system components are functioning properly Alarm...

Page 9: ...ory calibration parameters are also stored in nonvolatile memory The factory default settings are shown below Temperature mode C unlocked Site mode Ear Site text On The biotech mode is accessed by pre...

Page 10: ...8 Infrared Tympanic Electronic Thermometer Temperature mode Option Display C unlocked Locked C Locked F F unlocked C C F F Site mode Option Display Oral Core Axillary Rectal Ear ORL CORE AX REC EAR...

Page 11: ...and or smudges for proper operation A mild detergent may be added to the water Do not use cleaners such as Spray Nine Phisohex Hibiclens or Vesta Syde disinfectants as they may result in damage to the...

Page 12: ...meter equilibrate in the room for 20 minutes before using For all other system errors reset the thermometer by installing a probe cover If the system error does not clear send the thermometer to the f...

Page 13: ...ondensing Storage Temperature Range 25 C to 55 C 13 F to 131 F up to 95 RH non condensing If the unit is stored at extremes it is recommended that the unit be checked on the field calibration checker...

Page 14: ...ry to return a unit for repair please contact PSS World Medical customer information desk at 800 777 4908 Parts Listing To order repair parts contact our customer service center or your sales represen...

Page 15: ...abeling Comments or Questions about Lumeon Products 800 777 4908 Manufactured for PSS World Medical Inc 4345 Southpoint Blvd Jacksonville FL 32216 Made in the USA www myselectonline com One or more U...

Reviews: