Infrared Tympanic Electronic Thermometer

Section I — Introduction

This operator’s manual was prepared for the operator of the Infrared Tympanic Electronic Thermometer. This 

manual contains instructions for use, precautions, and available maintenance and service information. To achieve 

satisfactory results, the operator must read this manual thoroughly before attempting to use the thermometer.

Initial Set Up 

• Unpack the Infrared Tympanic Electronic Thermometer and examine it for structural integrity.
• If using the wall or cart mount options (sold separately) for the base unit, consult the installation sheet provided with the 

respective system.

• When first used, the Infrared Tympanic Electronic Thermometer will show the factory preset settings: ear mode 

(EAR) and the Celsius scale (°C).

Section II — General Information

The Infrared Tympanic Electronic Thermometer is a fast, accurate, and convenient clinical instrument for 

measuring patient temperatures. The Infrared Tympanic Electronic Thermometer is an ear canal thermometer 

with measurement site equivalence modes including oral, axillary, core, and rectal equivalent temperatures. The 

equivalence modes are explained further in Section V, Instructions for Use. 

Features 

• Very fast temperature acquisition that meets CEN and ASTM standards - see Section XI.
• Peak Select System™
• Temperature measurement range of 33.0°C to 42.0°C (91.4°F to 107.6°F).
• Once temperature has been acquired, the °C/°F button alternates the displayed temperature between °C and °F.
• Audible and visual indication of completed temperature acquisition.
• Disposable single use probe covers aid in the prevention of a cross contamination of infectious diseases.  
• Low and dead battery indicators.
• “Sleep” mode stores the last temperature and conserves battery life.
• 15, 30, 45, and 60 second pulse timer functions.
• The thermometer housing can be cleaned with common cleaning agents. See Section VII, Cleaning for instructions.
• Base unit protects the thermometer tip and stores probe covers for easy access.
• Easy to read LCD display with icons.
• Unit is designed for ambidextrous use.
• Improved probe design to allow ease of use on the pediatric population.

Section III — Important Precautions

• The thermometer is a precision optical instrument. For reliable operation, handle carefully and do not drop.
• Read the operator’s manual thoroughly before using the Infrared Tympanic Electronic Thermometer.
• Prior to use, ensure that the probe tip is clean and clear of any material.  If the thermometer tip becomes soiled, 

gently clean with a lens wipe or lint free swab. The tip should appear shiny and free of fingerprints and/or debris.

Summary of Contents for LUMEON 3069

Page 1: ...n Scale Dimensions Revisions Rev Description Date K PMS 110Yellow Meghan Little 3069_Masterpage Tympanic Thermometers 1 1 Proportion As Provided Final Final Draft 09 03 09 PMS 110Yellow K INFRARED TYM...

Page 2: ...5 Temperature Display Toggle 5 Sleep Mode 5 Pulse Timer Mode 6 Thermometer Display Icons and Alarms 6 Biotech Mode 7 Section VI Preventative Maintenance 9 Section VII Cleaning 9 Section VIII Battery...

Page 3: ...Features Very fast temperature acquisition that meets CEN and ASTM standards see Section XI Peak Select System Temperature measurement range of 33 0 C to 42 0 C 91 4 F to 107 6 F Once temperature has...

Page 4: ...mperatures during cold weather conditions allow the patient to equilibrate to room temperature before use Excessive eardrum scarring may cause lowered temperature readings Under normal conditions ear...

Page 5: ...Thermometer probe covers could jeopardize the accuracy of the Infrared Tympanic Electronic Thermometer Check your probe cover carton for the LumeonTM logo to be certain that your thermometer will deli...

Page 6: ...PSS World Medical Inc will result in erroneous temperature measurement In order to aid in infection prevention and control always install a new probe cover prior to taking a temperature The probe cov...

Page 7: ...age Temperature Recall After a temperature has been acquired the thermometer will enter off mode after approximately 10 seconds Recall the temperature by pressing and releasing the scan button or by p...

Page 8: ...talled or the batteries have been changed the thermometer will perform a system reset The thermometer will perform an internal test to verify that the system components are functioning properly Alarm...

Page 9: ...ory calibration parameters are also stored in nonvolatile memory The factory default settings are shown below Temperature mode C unlocked Site mode Ear Site text On The biotech mode is accessed by pre...

Page 10: ...8 Infrared Tympanic Electronic Thermometer Temperature mode Option Display C unlocked Locked C Locked F F unlocked C C F F Site mode Option Display Oral Core Axillary Rectal Ear ORL CORE AX REC EAR...

Page 11: ...and or smudges for proper operation A mild detergent may be added to the water Do not use cleaners such as Spray Nine Phisohex Hibiclens or Vesta Syde disinfectants as they may result in damage to the...

Page 12: ...meter equilibrate in the room for 20 minutes before using For all other system errors reset the thermometer by installing a probe cover If the system error does not clear send the thermometer to the f...

Page 13: ...ondensing Storage Temperature Range 25 C to 55 C 13 F to 131 F up to 95 RH non condensing If the unit is stored at extremes it is recommended that the unit be checked on the field calibration checker...

Page 14: ...ry to return a unit for repair please contact PSS World Medical customer information desk at 800 777 4908 Parts Listing To order repair parts contact our customer service center or your sales represen...

Page 15: ...abeling Comments or Questions about Lumeon Products 800 777 4908 Manufactured for PSS World Medical Inc 4345 Southpoint Blvd Jacksonville FL 32216 Made in the USA www myselectonline com One or more U...