User Manual
EN
SELECT button
By the SEL button, you can select the functions within
the ranges of the function rotary switch
button
Backlight turn on/off. Pressing this button the backlight
can be turned on / off. After 20 sec the backlight will
turn off automatically.
Clamp Lighting switch
Pressing this button the clamp lighting switch can be
turned on / off. After 20 sec the backlight will turn off
automatically.
M
easUreMents
Current Measurement
Set the function switch to 600A~, or to 20A~ position.
Press the trigger to open the jaw. Fully enclose one con-
ductor.
Read the measured current on the display.
Note: Ensure that the trigger is closed properly while
measuring current in order to get the correct readings.
Warning: Ensure that the probe leads are disconnected
from the Meter while measuring.
Low Pass Filter (LPF) Measurement
The LPF mode eliminates high frequency noise in volta-
ge and current measurements by means of a low-pass
filter. The Low Pass Filter mode offers a cut-off frequ-
ency of 1 KHz.
Set the function switch to 600A~ LPF, or to 20A~ LPF
position.
Press the trigger to open the jaw. Fully enclose the con-
ductor.
Read the measurement on the display.
Note: Ensure that the trigger is closed properly while
measuring current in order to get the correct readings.
Warning: Ensure that the probe leads are disconnected
from the Meter while measuring.
Voltage Measurement
Set the function switch to V position then press the SEL
button to select voltage.
DC – direct current, AC – alternating current.
Insert the red test lead into the INPUT terminal and the
black test lead into the COM terminal.
Connect the test leads in parallel to the circuit under
test.
Read the voltage measurement on the display.
Resistance Measurement
Set the function switch to Ω position.
Insert the red test lead into the INPUT terminal and the
black test lead into the COM terminal.
Touch the test probe tips across the circuit or compo-
nent under test.
Read the resistance on the display.
Warning: To avoid damage or electric shock, discon-
nect power to the circuit under test and discharge all
high-voltage capacitors before testing resistance!
Continuity Measurement
Set the function switch to Ω position.
Press the SEL button to select continuity
The disp-
lay icon will change.
Insert the black test lead into the COM, and the red test
lead into the INPUT terminal.
Connect the test leads to the circuit under test. If the
resistance is less than 30Ω, a tone will sound.
Note: you can also check open/closed circuits by con-
tinuity check.
Warning: To avoid damage or electric shock, discon-
nect power to the circuit under test and discharge all
high-voltage capacitors before testing continuity.
Diode Test
Set the function switch to Ω position.
Press the SEL button twice to select diode test
function.
Insert the red test lead into the INPUT terminal and the
black test lead into the COM terminal.
Touch the red probe to the anode side and the black test
lead to the cathode side of the diode being tested. The
approx. forward voltage drop of the diode will be disp-
layed. If the connection is reversed, the „OL” sign will ap-
pear on the screen.
Warning: To avoid damage or electric shock, disconnect
power to the circuit under test and discharge all high-vol-
tage capacitors before operating the Meter.
Temperature Measurement
Turn the rotary switch to ºC/°F position.
Insert the type K connector to the Meter.
Insert the red test lead into the INPUT terminal and the
black test lead into the COM terminal.
Press the SEL button to select the temperature measure
(ºC or °F.)
Touch the temperature probe tip to the device under test
or leave the temperature probe in the open air to measu-
re temperature.
Read the temperature measurement on the display.
Warning: To avoid electric shock, do not operate the
temperature probe in electrical environment and do not
measure in microwave. Be aware when measuring tem-
perature as the measured surface can burn you or can
cause injury because of the high temperature.
NCV (Non-Contact-Voltage) Detector
Turn the rotary switch to NCV position.
Place the Meter near to the cable. Audible beeping
and flashing LED will occur when a source of electrical
energy is detected. The sound and flashing will be more
frequent as soon as the electrical energy is closer to the
Meter.
Presence of phase detector function: Insert the red test
lead into the INPUT terminal then touch the probe tip
to the naked end of the cable. If phase is present, an
audible warning will sound and the LED light will flash.
Warning: without direct connection, the detector
cannot measure the correct range because the mea-
surements always depend on the thickness of the
insulation. This function can be used only AC voltage
measurements.