2
866.4.Maxtec www.maxtec.com
BAT
1.2 Symbol Guide
The following symbols and safety labels are found on the MaxVenturi:
1.3 Product Primary Functions
The primary function of the MaxVenturi is to deliver a mixed gas of oxygen and air
to either a heated humidifier system or directly to a patient. One of the primary
accessories to the MaxVenturi is a Fisher & Paykel heated humidifier (MR850)
and the Optiflow high flow patient interface system and Comfort Flo
®
Circuits
(2415 & 2416) with the Hudson RCI
®
Neptune
®
Heated Humidifier (425-00).
Several other patient interface circuits listed in section 2.5 can also be used
with the MaxVenturi.
The MaxVenturi flow meter is labeled to compensate for the back pressure by
the humidifier/patient interface system.
» The numbered scale on the flow meter that corresponds to the compensated flow rate
for the Fisher & Paykel Optiflow System and Hudson RCI
®
Neptune
®
Heated Humidifier.
» The lettered scale on the flow meter corresponds to the flow rates for the other patient
interface circuits listed in the chart in Section 2.5
Warning
On/off Button
Calibration Required
Room Air Inlet
Calibration Button
Low Battery
Percent
ETL Classified conforms
to ULSTD 60601-1
Do not throw away. Follow local
guidelines for disposal.
Patient Outlet
Adjustable Flow
CAL
%
EC REP
Authorized Representative
in the European Community
LOT
IPx1
SN
Lot code/Batch code
Drip Proof
Serial Number
REF
LPM
Catalog Number
Liter Per Minute Flow
Manufacturer
Follow instructions for use.
Do Not
Read Flow Meter Float
At Center
Federal Law (USA) restricts this device to
sale by or on the order of a physician
only
!