![Maxtec MaxVenturi R211P03 Operating Manual & Instructions For Use Download Page 17](http://html1.mh-extra.com/html/maxtec/maxventuri-r211p03/maxventuri-r211p03_operating-manual-and-instructions-for-use_1745656017.webp)
11
866.4.Maxtec www.maxtec.com
6.0 CHANGING THE OXYGEN SENSOR
NOTE:
For new sensor installation, follow procedure for new sensor installation in section 2.1.
7.0 CLEANING AND MAINTENANCE
Store the MaxVenturi at a temperature similar to that of its daily use.
7.1 Routine Maintenance
Maxtec recommends the performance check listed in the MaxVenturi
Preventive Maintenance Manual be conducted prior to placing the MaxVenturi
into clinical use and periodically thereafter. If the MaxVenturi does not function
as described in the performance check, discontinue use of the device pending
service by trained technicians or contact your Maxtec distributor or Maxtec at:
2305 South 1070 West Salt Lake City, Utah 84119.
The MaxVenturi's flow and O
2
adjustment valves should be replaced as needed.
The full MaxVenturi device should be overhauled and serviced a minimum of every 4
years with R211P30-001 and R211P30-002 to replace all O-rings, valves and knobs.
7.2 Device Cleaning
When cleaning or disinfecting the exterior of the MaxVenturi, take appropriate
care to prevent any solution from entering the device. Do not immerse
device in fluids. MaxVenturi external surfaces may be cleaned using a cloth
moistened with 65-70% isopropyl alcohol/water solution or a germicidal wipe.
7.3 Oxygen Sensor Cleaning
Clean the sensor with a cloth moistened with 65-70% isopropyl alcohol/water solution.
» Maxtec does not recommend use of spray disinfectants because they can contain
salts which can accumulate in the sensor membrane and impair readings.
» The oxygen sensor is not intended for steam, ethylene oxide or radiation sterilization.
2
3
1
Turn off analyzer.
Disconnect sensor from
sensor cable.
Remove sensor from
sensor port. Dispose of
sensor according to local
guidelines.