 WARNINGS

 

!

Indicates a potentially hazardous situation, if not avoided, could result in death or serious injury.

» This device is not intended for use with life supporting devices/systems.

»  Failure to comply with the warnings and precautions in this manual could result in 

instrument damage and possibly jeopardize the wellbeing of the patient and/or health care 

professional. Improper use of this device can cause inaccuracy of flow and oxygen readings 

which can lead to improper treatment, hypoxia or hyperoxia, or other patient injury or 

discomfort. 

Follow the procedures outlined in this user manual.

» Not for use in an MRI environment.
»   If the O

2

% drifts away from the level to which it was set, check to make sure the nasal prongs 

on the patient interface are not occluded by sputum or the nasal septum. Flow restriction to the 
circuit or patient interface will cause the oxygen level to increase. Flow restriction downstream of 
the venturi unit will not be detected by the flow meter.

» This device does not have any alarms for interruption in oxygen supply.
»  Allow oxygen reading to stabilize before adjusting oxygen content.
»  This device does not have any alarms for oxygen level high or low alarms.
»  Never allow excess length of tubing near the patient’s head or neck which could result in strangulation.
»  Use only Maxtec replacement sensors. Use of any other sensor will void warranty and may lead 

to product damage, product malfunction, improper treatment to patient, hypoxia, or hyperoxia.

 DO NOT

 

use this device near any type of flame or flammable/explosive substances, vapors or 

 atmosphere. Operating the oxygen analyzer in these environments may result in fire or explosion.

» This device in its entirety (including electronics) is not suitable for use in the presence of 

   flammable anaesthetic mixtures or in an atmosphere of explosive gases. Operating the oxygen 

   analyzer in these environments may result in fire or explosion.

 DO NOT

 attach a humidifier or any other source gas to the room air inlet. It should be occupied  

 at all times by the filter listed in the disposables list. The inlet filter prevents 

 entrainment of ambient contamination and silences venturi noise. 

The filter provided with the 

 MaxVenturi is single-patient use only.

   

» Use of this device with a pressurized oxygen bottle may result in inaccurate oxygen  
   concentration readings when used above 40 LPM and at high oxygen concentrations. High tank 
   pressures result in cooling oxygen supply temperatures which affect the accuracy of the oxygen 
   sensor. It is suggested that the device be connected with a long supply hose. Use a 15 ft supply 
   hose where possible — Maxtec P/N (R127P35).
» Use patient circuits that are approved for use with the manufacturer’s humidifier as listed in 
   their individual instructions for use.

 DO NOT

 attempt to clean the inside of the flow meter. If any malfunction is detected in the 

 function of the flow meter, if any debris or contamination is detected in the flow meter, or if the float 
 sticks in the flow tube, discontinue use immediately and return the device to Maxtec for service.

» Never install the sensor in any location other than the sensor port in the device.

SENSOR WARNINGS:

 The Maxtec MAX-250 oxygen sensor is a sealed device containing a mild 

acid electrolyte, lead (Pb), and lead acetate. Lead and lead acetate are hazardous waste constituents 
and should be disposed of properly, or returned to Maxtec for proper disposal or recovery.

 DO NOT

 use ethylene oxide sterilization. 

 DO NOT

 immerse the sensor in any cleaning solution, autoclave or expose the sensor to high 

 temperatures.

» Dropping or severely jarring the sensor after calibration may shift the calibration point enough to   
   require recalibration.

CAUTION:

 Indicates a potentially hazardous situation, if not avoided, could result in minor or 

moderate injury and property damage.
» The MaxVenturi is intended for use with specific patient interface configurations. The numbered scale    

   is intended for use with Fisher & Paykel* heated humidifier (MR850) and the Optiflow high flow patient   

   interface system (OPT846, OPT570) and Comfort Flo

®  

Circuits (2415 & 2416) with the Hudson RCI

®

  

   

Neptune

®

 

Heated Humidifier (425-00). The lettered scale (Labeled A through F) is intended for the 

   alternate patient circuits found on the chart provided in Section 2.

Summary of Contents for MaxVenturi R211P03

Page 1: ...O P E R A T I N G M A N U A L I N S T R U C T I O N S F O R U S E R211M03 Rev Q R 2 1 1 P 0 3 MaxVenturi A B C D E F 60 50 40 30 20 10 02...

Page 2: ......

Page 3: ...ars from the date of shipment from Maxtec Based on Maxtec s product evaluation Maxtec s sole obligation under the foregoing warranty is limited to making replacements repairs or issuing credit for equ...

Page 4: ...l times by the filter listed in the disposables list The inlet filter prevents entrainment of ambient contamination and silences venturi noise The filter provided with the MaxVenturi is single patient...

Page 5: ...ult in false readings from the flow meter The MaxVenturi is not intended for steam ethylene oxide or radiation sterilization DO NOT clean with ethanol or acetone After cleaning and before use on a pat...

Page 6: ...ent Interfaces 7 2 5 3 Inlet Filter 9 3 0 FACTORS INFLUENCING ACCURATE READINGS 9 3 1 Elevation Pressure Changes 9 3 2 Temperature Effects 9 4 0 CALIBRATION ERRORS AND ERROR CODES 9 5 0 CHANGING THE B...

Page 7: ...2 11 10 Flow meter Oxygen On Off Valve Inlet Oxygen Fitting Oxygen Control Knob O2 Flow Control Knob Room Air Inlet Room Air Inlet Filter ON OFF Button Calibration Button MAX 250 Series Oxygen Sensor...

Page 8: ...y the humidifier patient interface system The numbered scale on the flow meter that corresponds to the compensated flow rate for the Fisher Paykel Optiflow System and Hudson RCI Neptune Heated Humidif...

Page 9: ...displayed O2 percentage see Factors Influencing Accurate Readings in section 3 0 The MaxVenturi can be calibrated at 100 oxygen or room oxygen 20 9 The One Touch calibration will assume one of these t...

Page 10: ...ir 20 9 will result in an incorrect measurement of oxygen concentration 2 2 2 100 Oxygen Calibration 2 6 2 2 Minutes 1 2 3 4 5 6 1 5 Shut off Oxygen supply Remove sensor from port Wait two 2 minutesfo...

Page 11: ...7 8 2 Minutes Turn flow knob a few turns to flow gas Wait 2 minutes for sensor to equilibrate Press and hold CAL button for three 3 seconds Wait for display to read 100 Remove plug from room air inlet...

Page 12: ...ettings 4 5 1 2 3 1 2 3 1 2 3 1 2 3 1 3 Adjust flow to new desired flow rate Readjust oxygen concentration to desired setting and wait for analyzer to equilibrate Temporarily disconnect patient circui...

Page 13: ...ient Interfaces Optiflow large nasal cannula Fisher Paykel OPT544 Optiflow medium nasal cannula Fisher Paykel OPT846 Optiflow Tracheostomy direct connection Fisher Paykel OPT570 Pediatric aerosol mask...

Page 14: ...Comfort Flo Plus Cannula Large 2412 11 Hudson RCI Comfort Flo Plus Cannula Medium 2412 12 Hudson RCI Comfort Flo Plus Cannula Small 2412 13 F P Direct Connect OPT570 F P Large Nasal Cannula OPT846 F P...

Page 15: ...sensor to equilibrate to a new ambient temperature ATTENTION CAL Err St may result from a sensor that has not reached thermal equilibrium 4 0 CALIBRATION ERRORS AND ERROR CODES The analyzer has a sel...

Page 16: ...THE BATTERIES 1 2 3 4 5 6 7 8 9 Turn off analyzer Use screwdriver to remove four 4 screws in back of device Pull front cover off away from unit Turn on analyzer to verify power Reattach battery pack t...

Page 17: ...s needed The full MaxVenturi device should be overhauled and serviced a minimum of every 4 years with R211P30 001 and R211P30 002 to replace all O rings valves and knobs 7 2 Device Cleaning When clean...

Page 18: ...s 2 x 1 5 Volts Battery Life approximately 5000 hours with continuous use Low Battery Indication BAT icon displayed on LCD Sensor Type Maxtec MAX 250 series galvanic fuel cell Expected Sensor Life 1 5...

Page 19: ...d Operating Instructions R125P03 002 MAX 250E Oxygen Sensor RP34P02 Air inlet filter 9 2 Standard Replacement Parts and Accessories Part Number Item R127P35 15 coiled polyurethane oxygen hose with DIS...

Page 20: ...l knob completely closed Possible Cause It is very likely the device needs to be calibrated Also check to make sure the humidifier and patient delivery disposables are the correct size and installed c...