background image

ENGLISH 

10

 

WWW.MAXTEC.COM  •  (800) 748-5355

6.0 CLEANING AND MAINTENANCE

6.1 Cleaning

The external surfaces of the device and its accessories can be cleaned and disinfected using the 
process detailed below. Under normal use conditions, the surfaces of the sensor and T-adapter 
/ flow diverter that come in contact with gas delivered to the patient should not become 
contaminated. If you suspect that the sensing face of the sensor or internal surfaces of the 
T-adapter / flow diverter have become contaminated, these items should be discarded and 
replaced. Store the MaxO2 ME in a clean, dry location when not in use.

1.  The MaxO2 ME needs to be cleaned between each patient's use.
2.  Using Super Sani-Cloth germicidal disposable wipes (medical grade 2-in-1 cleaning / 

disinfecting wipes) remove all visible contamination from the external surfaces of the 
device and its accessories. Be sure to closely inspect and remove contamination from 
seams and recesses on the device that may trap contaminants.

3.  After all visible contamination is removed, use a second germicidal wipe to thoroughly 

wet the surfaces of the device and accessories. Allow to remain  wet for 4 minutes. Use 
additional wipes if needed to assure surfaces are wetted continuously for 4 minutes.

4.  Allow device to air dry.
5.  Visually inspect each component for visible contamination.

 CAUTION:

 Excessive rubbing of labels may cause them to become illegible.

 DO NOT 

spray cleaning solutions directly onto the monitor, sensor or buzzer opening.

 DO NOT 

immerse the MaxO2 ME or sensor into liquid decontamination agents.

 DO NOT 

use strong solvent cleaners.

 DO NOT 

allow cleaning liquids to contact the face of the sensor as this may impair the 

readings of the sensor.

 DO NOT 

attempt to sterilize the MaxO2 ME with steam, ethylene oxide or irradiation. 

6.2 Alarm Testing

Periodic testing of alarms should be performed on a yearly basis.

To check the low alarm, adjust the low alarm setting to 23% or higher and expose the sensor to 
room air (20.9%). The low alarm LED should flash with the alarm sound.

To check the high alarm, adjust the low alarm setting to 17% or lower and the high alarm 
setting to 18% and expose the sensor to room air (20.9%). The high alarm LED should flash 
with the alarm sound. If one or both alarms malfunction, contact Maxtec Certified Service 
Technician.

6.3 Replacing Sensor Cable

After extended use or abuse to the sensor cable, the cable may begin to wear and lose its 
ability to properly retract. 

The cable can be removed and replaced by disconnecting the threaded locking shroud at the 
sensor and monitor ends of the cable. Use only the Maxtec approved cable called out in Section 
9.0 Spare Parts and Accessories.

NOTE:

 Ensure the cable locking shroud is fully threaded on the sensor and the monitor.

7.0 SPECIFICATIONS

7.1 Base Unit Specifications

Measurement Range ............................................................................................................. 0.0-100%
Resolution .......................................................................................................................................0.1%
Accuracy and Linearity .......................................... ±1% of full scale at constant temperature, R.H.

and pressure when calibrated at full scale

Total Accuracy ...........................±3% Actual oxygen level over full operating temperature range
Response Time .........................................90% of final value in approximately 15 seconds at 23°C
Warm-up Time ............................................................................................................. none required
Operating Temperature ............................................................................. 15°C - 40°C (59°F - 104°F)
Storage Temperature .................................................................................. -15°C - 50°C (5°F - 122°F)
Atmospheric Pressure ................................................................................................ 800-1013 mBars 
Humidity .....................................................................................................0-95% (non-condensing)
Power Requirements ............................................................ 4, AA Alkaline batteries (4 X 1.5 Volts)
Battery Life ........................................................................approximately 5000 hours in typical use
Low Battery Indication ................................................................ "LOW BAT" icon displayed on LCD
Sensor Type ................................................................................ Maxtec Max-550E galvanic fuel cell
Expected Sensor Life .........................  >1,500,000% O2 Hours over 2 years in typical applications
Alarm System .............................................................. high/low alarms, flashing red/yellow LEDs, 

nominal 975Hz audio buzzer 

(according to IEC 60601-1-8 Audible Alarms in Medical Equipment)

Alarm Volume (all priorities) .............................................................. 70 dB(A) ± 7 dB(A) at 1 meter
Low Oxygen Alarm Range ....................................................15%-99% (>1% lower than high alarm)
High Oxygen Alarm Range ................................................ 16%-100% (>1% higher than low alarm)
Alarm Accuracy ................................................................................  exact to displayed alarm value
Dimensions ..................................................3.6"(W) x 5.8"(H) x 1.2"(D) [91mm x 147mm x 30mm]
Weight ............................................................................................  approximately 0.89 lbs. (.40 kg)
Cable Length ............................................................................................... 9 ft. (3m) fully extended
Diverter Fitting ...............................................................fits industry standard, 15 mm "T" adapter

8.0 APPLICATIONS

8.1 Exposure to Anesthetic Gases

Because of the unique chemistry of the oxygen sensors provided with the MaxO

2

 ME monitor, 

there are no significant effects when exposed to commonly used anesthetic gases, however, 
the monitor is not designed for exposure to flammable gas mixtures (See WARNING page 
2).

INTERFERENT

VOLUME % DRY

INTERFERENCE IN O2%

Nitrous Oxide

60% balance O2

<1.5%

Halothane

4%

<1.5%

Enflurane

5%

<1.5%

Isoflurane

5%

<1.5%

Helium

50%, balance O2

<1.5%

Sevoflurane

5%

<1.5%

Desflurane

15%

<1.5%

NOTE:

 Balance mixture 30% O2/70%N2O, unless otherwise specified.

8.2 Calibration Techniques in Pressurized 

Systems

Summary of Contents for MaxO2ME

Page 1: ...MaxO2 ME Instructions for Use E N G L I S H R230M01 001 REV H...

Page 2: ...60601 1 2formedicalproducts Theselimitsprovidereasonableprotectionagainstelectromag netic interference when operated in the intended use environments described in this manual MRI Notice Thisequipment...

Page 3: ...comes in con tact with infectious agents Gasleaksthatcauseroomairtomixwiththegassamplemaycauseinaccurateoxygenread ings EnsuretheO ringsonthesensorandflowdiverterareinplaceandintactpriortouse DONOTexp...

Page 4: ...g Oxygen Calibration 7 3 0 OPERATING INSTRUCTIONS 7 3 1 Alarm Setting Procedure 7 3 1 1 Low Alarm Setting 7 3 1 2 High Alarm Setting 8 3 1 3 Smart Alarm Mode 8 3 2 Basic Operation 8 3 3 Alarm Conditio...

Page 5: ...mponent Identification 1 LOWALARMLED Inalowalarmcondition theyellow LOWALARM LEDwillflashonce every two seconds accompanied by the audio buzzer If the Oxygen level is below 18 the red LOW ALARM LED wi...

Page 6: ...sensor after calibration may shift the calibra tion point enough to require recalibration CAUTION The flow diverter for the sensor is for use with flowing gases only DO NOT use the diverter when perf...

Page 7: ...1 Allow adequate time for the sensor to equilibrate to a new ambient temperature 2 When used in a breathing circuit place the sensor upstream of the heater 3 For best results perform the calibration...

Page 8: ...from the low alarm setting Pressing the Down key will subtract one from the high alarm setting and add one to the low alarm setting In other words the Up Arrow widens the alarm band and the down arro...

Page 9: ...the new sensor and tighten the cable connector 6 Screw the flow diverter onto the new sensor 7 Wait approximately 20 minutes for the sensor to reach equilibrium 8 Calibrate the new sensor NOTE If the...

Page 10: ...axtec Certified Service Technician 6 3 Replacing Sensor Cable After extended use or abuse to the sensor cable the cable may begin to wear and lose its ability to properly retract The cable can be remo...

Page 11: ...or the reading to stabilize before calibration Also note that the monitor may sense that the concentration is changing and an E07 error code will display Sensors come supplied with a flow diverter The...

Page 12: ...AC mains input lines IEC 61000 4 11 Dip 95 0 5 periods Dip 60 5 periods Dip 30 25 periods Dip 95 5 seconds Recommended separation distances between portable and mobile RF communications equipment and...

Page 13: ...os amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic si...

Page 14: ...2305 South 1070 West Salt Lake City Utah 84119 800 748 5355 www maxtec com...

Reviews: