Maxtec MaxO2ME Instructions For Use Manual Download Page 11

WWW.MAXTEC.COM  •  (800) 748-5355 

11

 ENGLISH

Similar to other oxygen sensors, the Maxtec MAX series sensors measure the partial pressure 
of oxygen in a gas stream. This is correlated to read “percent oxygen” on the MaxO2 ME moni-
tor. It is important to note that the sensor output is directly proportional to the partial pressure 
of oxygen. Thus, one must take into consideration the effect of exposing the sensor to various 
gas sample pressures.

For example, if a monitor is calibrated to read 20.9% in ambient air (atmospheric pressure) and 
then exposed to a pressurized gas sample containing a known concentration of oxygen, the 
monitor will display a reading greater than the actual oxygen percentage.

This is because the monitor was originally calibrated at atmospheric pressure (0 PSIG) then 
exposed to a higher pressure sample (i.e., 5 PSIG).

The greater the difference in pressure, the greater the difference in sensor signal (oxygen 
reading on the monitor).

If a monitor is calibrated on a pressurized gas sample containing a known concentration of 
oxygen and then exposed to ambient air (atmospheric pressure), the monitor will display a 
reading less than the actual oxygen percentage.To avoid confusion, the monitor can be cali-
brated at a single point on a gas stream similar to the application. If, for example, the purpose 
of the monitor is to measure oxygen in a concentrator or anesthesia application, the optimal 
results may be attained by calibrating the instrument on a gas of similar concentration and 
pressure. This would typically be done by connecting to a cylinder of a known high concentra-
tion of oxygen calibration gas and adjusting the flow and pressure to match the application 
before calibrating the instrument.

8.3 Calibration Errors

The MaxO2 ME monitor has a self test feature built into the software to detect faulty cali-
brations. During calibration, if the signal from the oxygen sensor is outside the limits stored 
within the instrument’s memory, a flashing E01 or E05 error code is displayed. The error code 
is displayed to indicate that either the sensor should be replaced or that there is a fault in the 
calibration process. A few simple hints can prevent calibration errors. If you try to calibrate the 
monitor before the reading has stabilized, the E01 or E05 error code may appear. For example, 
if the monitor had just been calibrated on a known high concentration of oxygen source gas 
and then exposed to ambient air, you should wait until the reading has stabilized. 

If you try to calibrate in room air before the sample line has cleared, the sensor may actually 
be exposed to residual oxygen. The signal from the sensor would still be high and considered 
out of range for air, thus resulting in an E05 or E07 error code. The proper procedure is to wait 
for the reading to stabilize before calibration.

Also note that the monitor may sense that the concentration is changing and an E07 error 
code will display.

Sensors come supplied with a flow diverter. The flow diverter helps direct the gas in a 
T-adapter up to the sensor for analysis. The flow diverter should be only used with a flowing 
gas. When using the sensor in a non-flowing environment, remove the diverter tip.

9.0 SPARE PARTS AND ACCESSORIES

PART NUMBER

ITEM

R140P02

Max-550E Sensor

R228P87

Battery Cover

R228P16

Sensor Cable

R228P10

Kickstand

R230M01

MaxO2 ME Operation Manual

R207P17

Barbed Concentrator Adapter for Sensor

PART NUMBER

ITEM

R205P86

Monitor/Analyzer Wall Mount Bracket

R206P75

Monitor/Analyzer Pole Mount Clamp

RP16P02

Maxtec Approved Tee Adapter (15mm I.D.)

R110P10-001

Sensor Flow Diverter

R230P10

Maxtec Approved External Power Supply

Repair of this equipment must be performed by a Maxtec Certified Service Technician experi-
enced in repair of portable hand held medical equipment. 
Equipment in need of repair should be sent to:
Maxtec
Service Department 
2305 South 1070 West
Salt Lake City, Ut 84119
1.800.748.5355
(Include RMA number issued by Customer Service)

10.0 ELECTROMAGNETIC 

COMPATIBILITY

The information contained in this section (such as separation distances) is in general specifi-
cally written with regard to the MaxO2 ME monitor. The numbers provided will not guarantee 
faultless operation but should provide reasonable assurance of such. This information may 
not be applicable to other medical electrical equipment; older equipment may be particularly 
susceptible to interference.

Note: Medical electrical equipment requires special precautions regarding electromagnetic 
compatibility (EMC) and needs to be installed and put into service according to the EMC infor-
mation provided in this document and the remainder of the instructions for use this device.

Portable and mobile RF communications equipment can affect medical electrical equipment.

Cables and accessories not specified within the instructions for use are not authorized. Using 
other cables and/or accessories may adversely impact safety, performance and electromag-
netic compatibility (increased emission and decreased immunity).

Care should be taken if the equipment is used adjacent to or stacked with other equipment; 
if adjacent or stacked use is inevitable, the equipment should be observed to verify normal 
operation in the configuration in which it will be used.

ELECTROMAGNETIC EMISSIONS

This equipment is intended for use in the electromagnetic environment specified below.
The user of this equipment should assure that it is used in such an environment.

EMISSIONS

COMPLIANCE 
ACCORDING TO

ELECTROMAGNETIC ENVIRONMENT

RF Emissions 
(CISPR 11)

Group 1

The MaxO2 ME uses RF energy only for its 
internal function. Therefore, its RF emissions 
are very low and are not likely to cause any 
interference in nearby electronic equipment.

Summary of Contents for MaxO2ME

Page 1: ...MaxO2 ME Instructions for Use E N G L I S H R230M01 001 REV H...

Page 2: ...60601 1 2formedicalproducts Theselimitsprovidereasonableprotectionagainstelectromag netic interference when operated in the intended use environments described in this manual MRI Notice Thisequipment...

Page 3: ...comes in con tact with infectious agents Gasleaksthatcauseroomairtomixwiththegassamplemaycauseinaccurateoxygenread ings EnsuretheO ringsonthesensorandflowdiverterareinplaceandintactpriortouse DONOTexp...

Page 4: ...g Oxygen Calibration 7 3 0 OPERATING INSTRUCTIONS 7 3 1 Alarm Setting Procedure 7 3 1 1 Low Alarm Setting 7 3 1 2 High Alarm Setting 8 3 1 3 Smart Alarm Mode 8 3 2 Basic Operation 8 3 3 Alarm Conditio...

Page 5: ...mponent Identification 1 LOWALARMLED Inalowalarmcondition theyellow LOWALARM LEDwillflashonce every two seconds accompanied by the audio buzzer If the Oxygen level is below 18 the red LOW ALARM LED wi...

Page 6: ...sensor after calibration may shift the calibra tion point enough to require recalibration CAUTION The flow diverter for the sensor is for use with flowing gases only DO NOT use the diverter when perf...

Page 7: ...1 Allow adequate time for the sensor to equilibrate to a new ambient temperature 2 When used in a breathing circuit place the sensor upstream of the heater 3 For best results perform the calibration...

Page 8: ...from the low alarm setting Pressing the Down key will subtract one from the high alarm setting and add one to the low alarm setting In other words the Up Arrow widens the alarm band and the down arro...

Page 9: ...the new sensor and tighten the cable connector 6 Screw the flow diverter onto the new sensor 7 Wait approximately 20 minutes for the sensor to reach equilibrium 8 Calibrate the new sensor NOTE If the...

Page 10: ...axtec Certified Service Technician 6 3 Replacing Sensor Cable After extended use or abuse to the sensor cable the cable may begin to wear and lose its ability to properly retract The cable can be remo...

Page 11: ...or the reading to stabilize before calibration Also note that the monitor may sense that the concentration is changing and an E07 error code will display Sensors come supplied with a flow diverter The...

Page 12: ...AC mains input lines IEC 61000 4 11 Dip 95 0 5 periods Dip 60 5 periods Dip 30 25 periods Dip 95 5 seconds Recommended separation distances between portable and mobile RF communications equipment and...

Page 13: ...os amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic si...

Page 14: ...2305 South 1070 West Salt Lake City Utah 84119 800 748 5355 www maxtec com...

Reviews: