background image

ENGLISH 

6

 

WWW.MAXTEC.COM  •  (800) 748-5355

OXYGEN CONCENTRATION —

 Current oxygen concentration percentage from the oxygen 

sensor.

HIGH ALARM INDICATOR —

 

 High alarm symbol used for identifying the high alarm 

setpoints and when a high alarm is triggered.

OXYGEN HIGH ALARM LIMIT —

 High oxygen alarm setpoint. Audible and visual alarms 

will trigger when this limit is exceeded. Double dashes (--) indicates the alarm is 
inactive.

CALIBRATION REMINDER —

 

 The calibration reminder symbol is located at the 

bottom of the display. This symbol will be lighted after one week has elapsed from the 
previous calibration.

SLEEP MODE INDICATOR —

   The sleep mode Indicator is used to reduce battery con-

sumption. See section 3.5 Sleep Mode Operation.

LOW BATTERY INDICATOR —

 

  The low battery indicator is located at the bottom 

of the display and is only activated when the voltage on the batteries are below a nor-
mal operating level and need to be replaced.

ALARM SILENCE/SMART ALARM INDICATOR —

 When the silent key 

 is pressed the 

indicator will display with cross bars 

 to alert condition. When Smart Alarm Mode 

Button is pressed the indicator will display with T-bars   to alert condition.

OXYGEN LOW ALARM LIMIT —

 Low oxygen alarm setpoint. Audible and visual alarms will 

trigger when this limit is exceeded.

LOW ALARM INDICATOR —

 

 Low alarm symbol used for identifying the low alarm 

setpoints and when a low alarm is triggered.

<18% ALARM INDICATOR —

 The <18% alarm indicator is located above the Low Alarm 

Indicator digits. When the low alarm setting is set below <18%, the indicator will flash 
each second to alert the operator of this special condition. See section 3.1.1 for setting 
this low alarm condition.

1.4 Max-550E Oxygen Sensor

The Max-550E is a galvanic, partial pressure sensor that is specific to oxygen. It consists of 
two electrodes (a cathode and an anode), a FEP membrane and an electrolyte. Oxygen dif-
fuses through the FEP membrane and immediately reacts electrochemically at a gold cathode. 
Concurrently, oxidation occurs electrochemically at a lead anode, generating an electrical cur-
rent and providing a voltage output. Electrodes are immersed in a unique gelled weak acid 
electrolyte which is responsible for the sensors long life and motion insensitive characteristic. 
Since the sensor is specific to oxygen, the current generated is proportional to the amount of 
oxygen present in the sample gas. When no oxygen is present, there is no electrochemical 
reaction and therefore, negligible current is produced. In this sense, the sensor is self-zeroing.

 CAUTION:

 The Max-550E oxygen sensor is a sealed device containing a mild acid electro-

lyte, lead (Pb), and lead acetate. Lead and lead acetate are hazardous waste constituents and 
should be disposed of properly, or returned to Maxtec for proper disposal or recovery.

 CAUTION:

 Dropping or severely jarring the sensor after calibration may shift the calibra-

tion point enough to require recalibration.

 CAUTION:

 The flow diverter for the sensor is for use with flowing gases only.

 DO NOT 

use the diverter when performing static sampling, such as in incubators, oxygen 

tents, oxygen hoods, etc.

2.0 SETUP PROCEDURE

2.1 Battery Installation/Replacement

All MaxO2 ME units are powered by four, AA, alkaline batteries (4 x 1.5 Volts) and are shipped 
without the batteries installed. The battery compartment is accessible from the back side of 
the unit. Batteries should be changed by qualified service personnel. Use only brand name 
batteries. Replace with four AA batteries and insert per orientation marked on the device.

When batteries are installed in the MaxO2 ME, the unit initiates a self diagnostic test. All seg-
ments of the LCD readout are turned on for approximately 2 seconds. The audio buzzer sounds 
and the high and low alarm LEDs are illuminated. When the diagnostic test is completed suc-
cessfully, the word "CAL" will display and then automatically initiate a calibration.

 WARNING:

 Battery replacement by inadequately trained personnel could result in a 

safety hazard. The MaxO2 ME will automatically perform a new calibration any time the bat-
teries are removed or replaced. Ensure that the sensor is exposed to either 20.9% oxygen 
(room air) or 100% oxygen when changing the batteries to avoid mis-calibration.

To install the batteries:

1.  Release the thumb screw by turning it counter-clockwise until it pops out.
2.  Install the four, AA, alkaline batteries (4 x 1.5 Volts) in the unit, observing the 

orientation shown on the plastic inside the compartment.

3.  Slide the battery compartment cover back onto the case. Press in on the thumb screw 

while turning it clockwise until it engages the thread in the enclosure. Turn until it is 
lightly tightened. 

 DO NOT

 over-tighten. 

 WARNING:

 Electrical shock or damage to the equipment may occur if an inappropriate 

external power supply is used. Maxtec recommends using only the Maxtec approved external 
power supply as listed in Section 9.0 Spare Parts and Accessories.

To protect the unit from potential leaky battery damage always remove batteries when the 
unit is going to be stored (not in use for 30 days or more) and replace dead batteries with 
recognized name brand AA Alkaline batteries.

2.2 Calibrating the MaxO2 ME Monitor
2.2.1 Before You Begin

A protective film covering the threaded sensor face must be removed; wait approximately 20 
minutes for the sensor to reach equilibrium. 

Next, the MaxO2 ME monitor should be calibrated. Thereafter, Maxtec recommends calibra-
tion on a weekly basis. However, more frequent calibration will not adversely affect product 
performance.

Calibration of the instrument should be performed when the temperature of the gas stream 
changes by more than 3 degrees Celsius.

Changes in barometric pressure can affect the oxygen reading. A 1% change in the barometric 
pressure results in an error of 1% of actual reading (Example: If you are reading a 50% oxygen 

u

i

f

g

d

s p

a o

y

Summary of Contents for MaxO2ME

Page 1: ...MaxO2 ME Instructions for Use E N G L I S H R230M01 001 REV H...

Page 2: ...60601 1 2formedicalproducts Theselimitsprovidereasonableprotectionagainstelectromag netic interference when operated in the intended use environments described in this manual MRI Notice Thisequipment...

Page 3: ...comes in con tact with infectious agents Gasleaksthatcauseroomairtomixwiththegassamplemaycauseinaccurateoxygenread ings EnsuretheO ringsonthesensorandflowdiverterareinplaceandintactpriortouse DONOTexp...

Page 4: ...g Oxygen Calibration 7 3 0 OPERATING INSTRUCTIONS 7 3 1 Alarm Setting Procedure 7 3 1 1 Low Alarm Setting 7 3 1 2 High Alarm Setting 8 3 1 3 Smart Alarm Mode 8 3 2 Basic Operation 8 3 3 Alarm Conditio...

Page 5: ...mponent Identification 1 LOWALARMLED Inalowalarmcondition theyellow LOWALARM LEDwillflashonce every two seconds accompanied by the audio buzzer If the Oxygen level is below 18 the red LOW ALARM LED wi...

Page 6: ...sensor after calibration may shift the calibra tion point enough to require recalibration CAUTION The flow diverter for the sensor is for use with flowing gases only DO NOT use the diverter when perf...

Page 7: ...1 Allow adequate time for the sensor to equilibrate to a new ambient temperature 2 When used in a breathing circuit place the sensor upstream of the heater 3 For best results perform the calibration...

Page 8: ...from the low alarm setting Pressing the Down key will subtract one from the high alarm setting and add one to the low alarm setting In other words the Up Arrow widens the alarm band and the down arro...

Page 9: ...the new sensor and tighten the cable connector 6 Screw the flow diverter onto the new sensor 7 Wait approximately 20 minutes for the sensor to reach equilibrium 8 Calibrate the new sensor NOTE If the...

Page 10: ...axtec Certified Service Technician 6 3 Replacing Sensor Cable After extended use or abuse to the sensor cable the cable may begin to wear and lose its ability to properly retract The cable can be remo...

Page 11: ...or the reading to stabilize before calibration Also note that the monitor may sense that the concentration is changing and an E07 error code will display Sensors come supplied with a flow diverter The...

Page 12: ...AC mains input lines IEC 61000 4 11 Dip 95 0 5 periods Dip 60 5 periods Dip 30 25 periods Dip 95 5 seconds Recommended separation distances between portable and mobile RF communications equipment and...

Page 13: ...os amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic si...

Page 14: ...2305 South 1070 West Salt Lake City Utah 84119 800 748 5355 www maxtec com...

Reviews: