Chapter 5 Operation Guide
Figure 5
(1) Power-on PD alarm, light-emitting PD alarm: If the two alarm lights are on,
restart the laser.
The restart sequence is:
1: first turn off the laser's key switch and turn off the air switch;
2: Power on , the order is to open the air switch first, then open the key switch;
If there is still an alarm, please contact the after-sales .
(2) Temperature alarm: Please check whether the fan is normal, whether it is
blocked by dust or other things, and there is a temperature alarm, turn off the
laser, and still have an alarm for cooling for a period of time, then contact the
after-sales ;
(3) Configuration error M Alarm: Once this alarm is available, restart the laser .If
still have an alarm, then contact the after-sales ;
(4) Laser enable fault: If there is such an alarm, the LASER_EN interface of
the DB25 interface is generally not connected, and it is checked whether the
interface is high level, must be high level, and the laser can work;