COMMISSIONING
5 - 3
DSE-17
www.maxcess.eu
MI 1064 1 C
Inbetr
iebna
hme.fm
Precondition:
– There must not be any material in the sensor field of view.
– The protection panes of the sensor must be clean.
Calibration
∙
Press the S2 key and hold it for 5seconds
LED H2 flashes (long on - short off)
The brightness values of the sensor are calibrated.
The bar graph represents the minimum and maximum level
of brightness values.
∙
Press key S2 and hold 0.5s to complete the calibration
process:
– If the calibration data is correct it will be saved.
LED H2 is lit for 3seconds before the sensor returns to its
regular operation state.
– If the calibration data is not correct, LED H2 flashes
quickly for 10seconds.
The sensor returns to its regular operation state without
saving the data from the calibration.
or
Press the S2 key and hold it for 3seconds to return to the
regular operation state
Calibration data will not be saved.