- 9 -
7-1. Setting Zero and Span
There are two methods, automatic and manual, to enter and set the zero and span values:
Use either method according to the situation.
7-1-1. Manual Entry and Setting Method
(a) Setting Zero
Check and record the value of the arithmetic processing result indicated under no-load
conditions (nothing is flowing inside). Then, enter that value into the zero setting in the
parameter setting window.
(b) Setting Span
Check and record the current value of the arithmetic processing result indicated under real-load
conditions (the subject material is flowing at the maximum flow rate). Then, enter that value into
the span setting in the parameter setting window.
Displays measurement results.
(Flow rate in % is shown.)
Displays arithmetic processing results.
(The product of
vibration amplitude V frequency Hz is shown).
Checks and records the value.
With the zero setting on display, enter the value checked in
the previous step.
The numeric character which is flickering can be changed.
Use [ ] to change the flickering position, [ ] to change the
number, and [STO] to store the entered value.
Press [ ], [ ],
and [STO].
Press [STO] three
Press [MOD].
Displays measurement results.
(Flow rate in % is shown.)
Displays arithmetic processing results.
(The product of
vibration amplitude V frequency Hz is shown).
Checks and records the value.
With the span setting on display, enter the value checked in
the previous step.
The numeric character which is flickering can be changed.
Use [ ] to change the flickering position, [ ] to change the
number, and [STO] to store the entered value.
Press [STO]
three times.
Press [ ], [ ],
and [STO].
Press [MOD] twice.