6
Getting started
30
6-3
CO-U32m, Standard Digital Interface
Control a RK(400W/800W/1200W)/REK/REKJ series power supply connected to CO-U32m via
USB.
①
Set unit number
Set the UNIT number to “0”.
See section ”
4-2
Setting of UNIT number (unit with Standard Digital Interface)”.
②
Connect to USB
Connect USB connector of CO-U32m and USB controller(e.g.PC) with USB cable(available at
stores).
③
Connect RK(400W/800W/1200W)/REK/REKJ series power supply to CO-U32m
Connect CO-M cable connector(OUT) of CO-U32m to CO-M cable connector(IN) with CO-M
cable enclosed together with power supply.
④
Turn on the power of CO-U32m
Turn on POWER ON/OFF switch of CO-U32m.
⑤
Turn on the power of RK(400W/800W/1200W)/REK/REKJ series power supply
Turn on POWER ON/OFF switch of RK(400W/800W/1200W)/REK/REKJ series power supply.
⑥
Enable remote control
Send a command of “#0 REN” from USB controller.
REMOTE LED shall light on.
⑦
Enable the output
Send a command of “#0 ICN 100”, “#0 OVP100” ,”#0 SW1” from USB controller.
OUTPUT LED shall light on.
⑧
Set output voltage and output
Send a command of “#0 VCN xxxx” from USB controller. Enter numbers of 0 to 100 for xxxx.
Refer to Command list for details.
⑨
Stop the output
Send a command of “#0 SW0” from USB controller.
OUTPUT LED shall turn off.
⑩
Ending the operation
Turn off POWER ON/OFF switch of RK(400W/800W/1200W)/REK/REKJ series power supply.
⑪
Stop remote control
Turn off POWER ON/OFF switch of CO-U32m.
CAUTION
・
Voltage is output actually.
・
Do not connect any load to power supply before this procedure is
completed.
!