background image

Integrity Bath 

 

Model MB-80

 

Maintenance Manual 

  

07-06-15 
Page 15 of 16

 

2071 14

th

 Ave – Columbus, NE 68601 

800-798-5867  FAX: 402-563-9102 
www.mastercarebath.com

 

SCALE CALIBRATION 

Memo:  Before  calibrating  acquire  a  calibrated  25  lb. 
weight.  Note:  The  weight  must  calibrated  or  scale  will 

display  inaccurate  weights.  Chair  must  be  on  carder  with 
scale in order to calibrate. 

1.

 

Turn  the  scale  on  by  pushing  the  ON/ZERO  BUTTON. 
With  scale  on  ensure  the  lbs  display  is  showing.  If  not 
push the LBS/KG BUTTON until lbs. shows on the ·display. 
After the lbs. display is activated allow it to shut off on its 
own  

2.

 

Turn  the  scale  back  on  again  by  pushing  the  ON/ZERO 
BUTTON.  After  the  scale  turns  back  on  push  the 
ON/ZERO BUTTON and continue to hold down while also 
pushing the LBS/KG BUTTON down CAL will be displayed. 
Hold  both  buttons  down  for  about  3  to  4  seconds  until 
the  display  reaches  C  -  0  (zero)  then  release  both 
buttons. 

3.

 

Push  ON/ZERO  BUTTON  down  one  more  time  then 
release. The display will start counting down from 16 to 
O (zero). After the display reaches O (zero) a new display 
will  read  C-25.  Place  the  25  lb.  weight  centered  on  the 

chair. Again, push the ON/ZERO DOWN then release and 
the display will start counting down from 16 to 0. After 
the display reaches 0 a new display will read 25. Remove 
the 25lb weight. Your Scale is now calibrated. 

 

 

Summary of Contents for MB-80

Page 1: ...Integrity Bath Model MB 80 Maintenance Manual For All Maintenance Procedures Follow the Directions in this Manual...

Page 2: ...o their application and the durability requirements of the healthcare industry It is recommended that the bathing system be inspected and preventative maintenance performed using the following schedul...

Page 3: ...or and swing out of way Remove SOCKET HEAD CAP SCREWS from ROLLER BRACKET Save SS SPACERS between BRACKET and DOOR HINGE MOUNTING BASE b Add SS SPACER to each spacer group Reinsert SOCKET HEAD CAP SCR...

Page 4: ...STIC WEDGE index is on top of the PLASTIC BUSHING b Position ALUMINUM MOUNTING BLOCK to insure even engagement of PLASTIC WEDGE Check the clearance of the bottom edge of the DOOR See drawing below You...

Page 5: ...the SLIDE END CAP item 3 in drawing 803966 It should not be worn through on the side that contacts the outer surface of the door To replace 1 Remove old end cap 2 Clean any residual adhesive 3 Smear a...

Page 6: ...bove directions Insert special tool in drain basket Item 12 slots Clamp on center of special tool with a pair of Vice Grips Remove by turning counter clockwise Clean old seal material away from drain...

Page 7: ...itch Item 72 5 Disconnect the air tubing Item 79 from the switch nipple 6 Remove spade connectors from switch terminals noting which was connected to which terminal 7 Remove nut holding the switch in...

Page 8: ...the replacing the cartridge 2 Using a utility knife gently pry off the cap 6 that is located on top of the thermostatic control valve handle 7 3 Using a Philips screwdriver remove the screw 8 and the...

Page 9: ...0 798 5867 FAX 402 563 9102 www mastercarebath com GEN II MIXING VALVE DIAGRAM ITEM NO DESCRIPTION QTY 1 THERMOSTATIC VALVE 1 2 CARTRIDGE REPLACE THERMOSTATIC 1 3 O RING 3 4 HUB 1 5 SETTING RING 1 6 C...

Page 10: ...d or cracked replace Check aeration body tightness against tub floor If loose tighten until 0 ring part 36000 under the shoulder seals With tub full of water and air on Check area around aeration gask...

Page 11: ...Integrity Bath Model MB 80 Maintenance Manual 07 06 15 Page 11 of 16 2071 14 th Ave Columbus NE 68601 800 798 5867 FAX 402 563 9102 www mastercarebath com VARIABLE SPEED BLOWER ENTREE 500105...

Page 12: ...with the pump operation will turn on when the pump operation is activated The middle Red LED is for Bath Oil The bottom Red LEO is for Shampoo Priming Operation Remove the empty one gallon container o...

Page 13: ...ate Is the check valve inside the metering knob clogged To fix remove four screws to disassemble Inspect and clean ball spring and 0 ting Lubricate with liquid soap and reassemble Check the orifice to...

Page 14: ...es but doesn t stay positioned when releasing lever Rotate clevis mount counter clockwise one turn and reinstall Rotate another turn if chair still doesn t lock in place Replacement 1 Support chair ba...

Page 15: ...w it to shut off on its own 2 Turn the scale back on again by pushing the ON ZERO BUTTON After the scale turns back on push the ON ZERO BUTTON and continue to hold down while also pushing the LBS KG B...

Page 16: ...ng Assembly Door Handle Assembly 301176 804220 Console Plumbing Assembly 804221 Console Plumbing Pre assembly 804225 Reservoir Assembly 804245 Drain Shoe Assembly Integrity Reservoir 804255 Plumbing A...