background image

 

Filling the UV Unit with Water: 

 

NOTE: Plug the 1” normally closed solenoid into a dedicated 

electrical supply. This is only during the filling and disinfection 

of the piping!   

 

1.  Open the inlet ball valve about 25% to fill the UV vessel slowly.  

2.  Open the inlet and outlet ball valve to their 

full

 open position.      

3.  Open cold water faucet in kitchen and run until all air is purged. 

     Shut off faucet. 

4.  Inspect for leaks and repair if any are found.

 

 

Disinfecting the Piping System with Chlorine: 

 

NOTE

The MWC-EM meter will drop from a 100% and start to 

alarm once the chlorine is induced into the system! This is 

common and is caused by the reaction when chlorine and 

water are mixed and the UV meter eye senses interference in 

the UV vessel.  

 

1.

  Shut off the water to the inlet and outlet of the cartridge filter.   

2.

  Remove the cartridge sump and empty all water.  Fill with 

household chlorine (Clorox, non-scented), 5.25% sodium 

hypochlorite. 

3.

  Re-install the sump to the filter assembly and open the inlet and 

outlet valves. 

4.

  Open the cold water faucet at the furthest point in the piping until 

you detect chlorine odor.  Once the odor is detected, close that 

faucet.  Continue this procedure until the chlorine odor has been 

detected at every faucet.  If chlorine odor is not detected at all 

faucets, repeat steps 1 through 4. 

5.

  Close ball valve on outlet of cartridge filter. 

6.

  Allow the water to stand for at least (4) hours to kill the bacteria 

residual in the piping system.  

7.

  Unplug solenoid from outlet and plug into the lower right side of 

the MWC-EM Meter Power Supply. 

  

            ** 

No water should be used during disinfection. **

 

 

Summary of Contents for MWC-EM Series

Page 1: ...MWC EM Series Ultraviolet Sterilizers MWC EM7 MWC EM10 MWC EM20 Installation and Operation Manual 2022 VERSION...

Page 2: ...ts MWC Meter Power Supply Meter Power Supply Installation 7 Grounding Grounding Instructions 8 Quartz and O ring Installing Quartz and O ring Germicidal Lamp Installing Germicidal Lamp MWC EM Meter St...

Page 3: ...of our MWC EM Water Sterilizers Shipping Description The UV unit is shipped complete as follows The large box contains the UV vessel and inside that box you will also find the following 1 A box conta...

Page 4: ...the maximum allowable levels specified this will result in bacteria passing through the UV unit and into the piping system Unit Location 1 The MWC EM unit must be installed after the well tank and any...

Page 5: ...ng the quartz tube and germicidal lamp For proper clearance see Specification Table on page 2 6 Secure the bottom bracket to the wall by using three screws or fasteners not provided 7 Use quality unio...

Page 6: ...ring this is on the quartz eyepiece 4 Insert the UV sensor assembly with gray gland nut Hand tighten only 5 Attach the other end of the UV sensors black amp electrical connection to the lower right s...

Page 7: ...power 24 hours per day We recommend installing a surge protector to protect unit from power surges which are not covered by warranty DO NOT UNDER ANY CIRCUMSTANCE WIRE THE UV UNIT TO THE WELL PUMP SW...

Page 8: ...rly grounded The vessel must be grounded to the MWC Power Supply to make a complete ground MWC SERIES Grounding Lug Installation Grounding Lug Front View of Bracket Insert stripped green ground wire i...

Page 9: ...3 Gently insert the germicidal lamp into the quartz sleeve through the center of the clear gland nut located at the top of the vessel Be careful not to drop the lamp inside the vessel This could break...

Page 10: ...ses interference in the UV vessel 1 Shut off the water to the inlet and outlet of the cartridge filter 2 Remove the cartridge sump and empty all water Fill with household chlorine Clorox non scented 5...

Page 11: ...tlet with power 24 hours per day 3 Check all piping for leaks repair if necessary 4 Unplug the 1 normally closed solenoid from the electrical outlet and plug the solenoid into the lower right side of...

Page 12: ...driver Remove the black cap located on the upper right side of the meter 5 With the unit on and operating turn the lower screw clock wise turn at time until you reach 100 The MWC EM unit is now workin...

Page 13: ...y pulling up on the gray harness wire and discard the lamp Remember the lamp is glass and can cause injury to you if the lamp is dropped 5 Open the ball valve test port on outlet side of UV unit to de...

Page 14: ...port and plunger Defective solenoid replace Symptom Alarm is on and no flow of water and Meter is reading less than 50 and Lamp is on and Gland Nut is glowing blue Cause Solution Lamp is over a year o...

Page 15: ...where the ultraviolet sterilizer has been operated at pressure in excess of 100 pounds per square inch or at a temperature in excess of 100 degrees F or below 32 degrees F Incidental costs or conseque...

Page 16: ...driver Remove the black cap located on the upper right side of the meter 5 With the unit on and operating turn the lower screw clock wise turn at time until you reach 100 The MWC EM unit is now workin...

Page 17: ...nnection to the lower right side of the meter Insert the amp into the twist lock connector The connector will only fit into the connector one way Hand Tighten Only UV Sensor Assembly Quartz Dome Eyepi...

Page 18: ...nt is required before installation of the UV unit Water must meet criteria below Untreated Water Quality Maximum Allowable Level Chloride 250 00 ppm Color 0 00 units Hardness 8 00 gpg Iron 0 30 ppm Ma...

Page 19: ...ore consumption of water an EPA Lab Test is suggested MWC EM SERIES TROUBLESHOOTING Symptom No flow of water no Alarm and Gland Nut not glowing Cause Solution Electric Supply Check electrical outlet a...

Reviews: