Non-Contact Voltage (NCV)

1.Hold the “

NCV

” button.

2.Move the tip of the meter toward the object to be tested.
3.If the voltage detected is greater than 90V AC rms, the 
    NCV indicator will flash and the buzzer will sound.

Note:
• 

Even if no indication is given, voltage may still be present.  

  Do not rely solely on NCV detection to determine the 
  presence of voltage.

• 

When measuring AC/DC voltage, the NCV indicator may 

  flash due to induced voltage.

• 

External power sources/interference may trigger the 

  NCV indicator.

Temperature (Type-K Thermocouple)

1.nsert the “- COM“ end of the thermocouple into the 
   “

COM”

 jack and “V +” end into the “

INPUT” 

jack.

2.Turn the rotary switch to the “

TEMP

” position. The 

    default mode is °F; press “

FUNC

” to switch to °C if 

    necessary; press “

FUNC

” to switch back to °F.

3.Touch the tip of the thermocouple to the object to be 
    tested.
4.Read the measured temperature value on the display. 

Specifications

The meter should be calibrated annually between 
18

 ~ 28

 and a relative humidity less than 75%.

General Specifications

• Manual and auto range
• Power overload protection
• Maximum voltage between circuit and ground: 
  1000V  AC/DC
• Maximum working height: 2000m
• Display: LCD
• Maximum display value: 5999
• Polarity indication” automatically displays“-”
• Overload indication: “0L” or “-0L”
• Sampling frequency: 3 times/sec
• Units displayed: functions with units.
• Auto power off: 15 mins.
• Power : 1x 9V 6F22 battery
• Low battery indication: display shows“        ”
• Operating temperature: 0

 ~40

• Storage temperature:-10

 ~50

• Temperature coefficient: less than 0.1% accuracy/°C
• Size: 201×65×43mm
• Weight: 265g (Including batteries)

Summary of Contents for MS2601

Page 1: ...MS2601 Electrical Tester User s Manual CAT III 600 V 600V CAT III Hz A V Loz A Temp FUNC RAN Hz NCV MS2601 ELECTRICALTESTER 600V CAT III COM MAX 200A Temp A...

Page 2: ...ode 7 Backlight and worklight 8 Auto Power Off 8 Measurement Preparation 8 AC Current 9 Switching Between Functions 7 AC DC Voltage 10 Resistance 11 Diodes 12 Continuity 11 Capacitance 12 Non Contact...

Page 3: ...ument to Mastech no service will be provided without a return authorization The user is responsible for properly packing the unit and charges such as shipping freight and insurance charges The extent...

Page 4: ...e circuit first before switching between functions 12 Disconnect power and discharge all capacitors before measuring resistance capacitance continuity or diodes 13 Do not measure capacitance before th...

Page 5: ...een functions RAN button to switch to manual range button to hold the reading or to turn on backlight Hz button to switch between frequency and duty cycle NCV button to activate non contact voltage de...

Page 6: ...t automatically turns off To turn off the backlight manually hold again When the meter is in current mode turning on the backlight will also activate the worklight making it easier to use the clamp in...

Page 7: ...into INPUT jack 2 Turn the rotary switch to the V The default mode is DC voltage press FUNC to switch to AC voltage if necessary press FUNC to switch back to DC voltage 3 Connect the test leads across...

Page 8: ...istance is less than 50 the meter s buzzer will sound Note If the resistance is higher than 60M or the leads are disconnected the display shows OL Capacitance TO AVOID ELECTRICAL SHOCK AND INJURY TURN...

Page 9: ...lead into INPUT jack 2 Turn the rotary switch to the position Press FUNC to switch to AC voltage 3 Press Hz button to measure frequency 4 Press Hz again to measure duty if necessary 5 Connect the test...

Page 10: ...ault mode is F press FUNC to switch to C if necessary press FUNC to switch back to F 3 Touch the tip of the thermocouple to the object to be tested 4 Read the measured temperature value on the display...

Page 11: ...true reading will be shown 600mV 0 1mV 1000 A AC Voltage 0 8 reading 3 digits Range Resolution Accuracy 6V 0 001V 60V 0 01V 600V 0 1V Input impedance 10M Maximum input voltage 600V DC AC rms Frequency...

Page 12: ...measured resistance is less than 50 0 1 Description Function Resolution Continuity Over voltage protection 250V DC AC rms Diodes Displays forward biased voltage drop 0 001V Description Function Resolu...

Page 13: ...meter or personal injury To avoid electrical shock remove test leads from input jacks before cleaning To avoid false readings that could lead to possible electric shock replace the batteries as soon...

Page 14: ...electric shock make sure the probes are disconnected from the measured circuit before removingthe rear cover Make sure the rear cover is tightly screwed before using the instrument Warning Out of the...

Reviews: