Chapter 7: Alarms and Messages
Messages and Indications
The table below lists the types of messages that can appear on Root when using SedLine.
Error Message
Indication
SedLine is
Disconnected
Indicates that the SedLine module is not connected.
No Sensor
Connected
Indicates the sensor is not properly connected to the patient cable or the
electrodes of the sensor are not connected to the patient's forehead.
Incompatible
Sensor
Indicates the sensor type cannot be used in conjunction with SedLine.
Sensor Off Patient
Indicates the electrodes of the sensor are not connected to the patient's
forehead.
High Impedance
Indicates the impedance values of the sensor electrodes are too high.
Gel Bridging
Detected
Indicates that the active (L1, R1) and ground (CB) electrodes may have
gel between them.
Alarms and Indications
SedLine has a Patient State Index (PSi) audible and visual alarm.
Alarm
Text
Indication
Action
PSi High
> ##
Indicates the PSi is
greater than the high
alarm limit value
configuration setting.
If you wish to change the audible alarm limit: press
the red alarm bell on the top left of the window.
Note
: If you adjust the alarm limit, the selected value
will remain until it is adjusted by the user.
PSi Low
< ##
Indicates the PSi is less
than the low alarm limit
value configuration
setting.
If you wish to change the audible alarm limit: press
the red alarm bell on the top left of the window.
Note
: If you adjust the alarm limit, the selected value
will remain until it is adjusted by the user.