Chapter 5: Menu Options
Parameter Settings
Parameter configuration settings provide the user access to seven parameters: PSi, DSA,
SEFL, SEFR, EMG, SR, and ARTF.
To access a specific parameter's configuration settings
•
Press the parameter desired directly from the SedLine window.
To access all parameter configuration settings
1.
Press the Main Menu icon at the bottom right corner of the screen to access menu
options.
2.
From the Main Menu, press the SedLine icon.
3.
Press the Parameter Settings icon.
4.
Press the parameter that you want to configure.
All changes to configuration settings must be accepted by pressing
OK
when prompted. To
decline changes made, press
Cancel
.
PSi
About
An informational read-only screen with definition of PSi:
The Patient State Index (PSi) is related to the effect of certain anesthetic agents on a patient.
Alarms
Options
Description
Alarm
Priority
Factory
Default
Selectable
Settings
High Limit
Upper limit that triggers an
alarm
Low/Medium* 50
5 to 99 in steps
of 1, or Off
Low Limit
Lower limit that triggers an
alarm
Low/Medium* 25
Off, or 1 to 95 in
steps of 1
High Caution
Range
Value below High Limit to
trigger caution light
NA
Off
Off, or 1 to 10 in
steps of 1
Low Caution
Range
Value above Low Limit to
trigger caution light
NA
Off
Off, or 1 to 10 in
steps of 1