Radical Signal Extraction Pulse Oximeter Operator’s Manual
9-9
9
s e r v i c e a n d m a i n t e n a n c e
S e r v i c e a n d r e p a i r
REPAIR POLICY
Masimo or an authorized Service Department must perform warranty repair and service.
Do not use malfunctioning equipment. Have the unit repaired.
Please clean contaminated and/or dirty equipment before returning, following the
cleaning procedure described in Section 9,
Cleaning. Make sure it is fully dry before
packing the equipment.
To return the Radical unit for service, please follow the Return Procedure.
RETURN PROCEDURE
Please clean contaminated/dirty equipment before returning and make sure it is fully dry
before packing the equipment. Call Masimo at 800-326-4890 and ask for Technical Support.
Ask for an RMA number. Package the equipment securely – in the original shipping
container if possible – and enclose or include the following information and items:
■
A letter describing in detail any difficulties experienced with the Pulse Oximeter.
Please include the RMA number in the letter.
■
Warranty information – a copy of the invoice or other applicable documentation
must be included.
■
Purchase order number to cover repair if the oximeter is not under warranty, or for
tracking purposes if it is.
■
Ship-to and bill-to information.
■
Person (name, telephone/Telex/fax number, and country) to contact for any
questions about the repairs.
■
A certificate stating the oximeter has been decontaminated for bloodborne pathogens.
Return the Radical to the following shipping address:
WARNING:
DO NOT REMOVE THE COVER OF THE MONITOR EXCEPT FOR
BATTERY REPLACEMENT. AN OPERATOR MAY ONLY
PERFORM MAINTENANCE PROCEDURES SPECIFICALLY
DESCRIBED IN THIS MANUAL. REFER SERVICING TO
QUALIFIED SERVICE PERSONNEL TRAINED IN THE REPAIR OF
THIS EQUIPMENT.
Analog Output Test
1.
Disconnect all patient cables and sensors from the Radical. Turn the Radical off and
then on again.
2.
Connect the common lead of a digital voltmeter to the pin 2 (Ground) of the analog
output connector on the Radical. Connect the positive lead of the voltmeter to pin 9
(Analog 1) of the analog output connector.
3.
Enter the menu system and set the "Output", "Analog 1" to "0V Signal".Verify that the
voltmeter measures a voltage of approximately 0V.
4.
Enter the menu system and set the "Output", "Analog 1" to "1V Signal".Verify that the
voltmeter measures a voltage of approximately 1.0V.
5.
Repeat Steps 3 and 4, with the positive lead of the voltmeter connected to pin 15
(Analog 2).
6.
Connect a patient cable and sensor and verify that the voltage on pins 9 and 15 are
between 0V and 1.0V while measuring a saturation and pulse rate..
Battery Test
1.
Fully charge the Radical by placing the Handheld into the Docking Station and
connecting the AC power.
2.
Verify that the green Handheld Battery Indicator LED is lit up.
3.
When the Radical is fully charged the green Handheld Battery Indicator turns off.
4.
Turn the Radical on and verify that the Battery indicator shows a full charge.
s e r v i c e a n d m a i n t e n a n c e
9-8
Radical Signal Extraction Pulse Oximeter Operator’s Manual
9
For USA:
For Europe:
For Asia Pacific:
Masimo Corporation
40 Parker
Irvine, California 92618
949-297-7000
Masimo Europe Limited
304 RN6, Le Bois des Cotes 2
69760 Limonest
France
Tel: +33 (0) 472 17 93 70
FAX: +33 (0) 478 35 78 08
Masimo Japan Corporation
World Times Bldg. 4F
10-7, Ichiban-cho, Chiyoda-ku,
Tokyo 102-0082 JAPAN
Tel: 03 3237 3057
FAX: 03 3238 1110