Radical Signal Extraction Pulse Oximeter Operator’s Manual
3-5
3
s e t u p
SATSHARE SETUP
1.
Select the SatShare cable that is appropriate for the multiparameter monitor that is
being connected. Check the Masimo web site at www.masimo.com for the latest list
of available SatShare cables and validated instruments.
2.
Connect the labeled end of the cable to the SatShare Cable Connector port on the
back of the Docking Station. Tighten the connector screws for a secure connection.
3.
Connect the other end of the SatShare cable either to the sensor connector of the
multiparameter monitor’s SpO
2
cable or directly to the SpO
2
connector on the monitor.
4.
Verify that the Radical recognizes the correct cable. The name of the SatShare cable
will be displayed on the LCD screen when the SatShare mode is functional.
5.
Set the multiparameter monitor’s high and low saturation and pulse rate alarm limits
as appropriate.
6.
Set the multiparameter monitor's averaging time to the lowest setting (i.e. fastest
response). The Radical's ideal waveform obviates the need for additional averaging
by the monitor. If the multiparameter monitor's averaging time is not changed, the
time to display physiological changes in saturation on the monitor will be increased
with SatShare. However, the delay can be minimized by reducing the multiparameter
monitor's averaging time.
7.
While in the SatShare mode, if there are any significant discrepancies between the
readings from the Radical and those on the monitor displaying the values obtained
from SatShare, the values reported by the Radical are to be considered the correct
values.
8.
To use the Radical with SatShare while it is not connected to AC power, set the Power
Save parameter in the General menu to "No", refer to Section 4,
Operation. Please
note that if the Radical is used in this mode, the length of time the Radical can
operate on battery power will be significantly diminished.
9.
Set the SatShare Numbers and the Interface Alarms parameters in the General
menu according to Customer preference.
10.
If displaying the simulated waveform is not desirable, it is recommended to turn off
the pleth waveform display of the multiparameter patient monitor.
ADULT/NEO (HOSPITAL DEFINED) SETTINGS
The following table outlines settings that may be defined by the hospital. If enabled (via
a password protected screen), these settings will return to pre-defined values after a
power cycle. This mode is indicated by “Mode Adult” or “Mode Neo” on the Alarms
menu. See Section 4,
Output, for details on enabling and setting these parameters.
S a t s h a r e s e t u p
The Radical Pulse Oximeter has been proven to be accurate during patient motion and
low perfusion conditions. Saturation and pulse rate values from the Radical may be
displayed on a multiparameter monitor through the SatShare feature.
The SatShare feature provides an ideal, simulated waveform corresponding to the measured
saturation and pulse rate values determined by the Masimo SET technology. This
waveform may be used to display these values on multiparameter monitors through the
multiparameter monitor’s oximetry sensor or input connector.
It is recommended that the Radical is positioned close to the multiparameter monitor with
the Radical screen visibly displaying the plethysmographic waveform and the saturation
and pulse rate measurements.
CAUTION:
SIMULTANEOUS USE OF SATSHARE AND SERIAL PORT IS NOT
SUPPORTED.
ADULT / NEO SETTING
OPTION
CUSTOM SETTING
(PRE DEFINED)
SpO
2
HIGH/LOW
ALARM LIMIT
Set to pre-power down setting
Set to pre-defined
Adult or Neo setting
PULSE RATE
HIGH/LOW ALARM
LIMIT
Set to pre-power down setting
ALARM SILENCE
Set to 120 seconds
ALARM VOLUME
Set to pre-power down setting
ALARM DELAY
The 0 and 5 second setting is
retained, the 10 second setting is
set to 5 seconds
OPTION
DEFAULT SETTING
PULSE BEEP VOLUME
Set to pre-power down setting
TREND DISPLAY PARAMETERS
Set to pre-power down setting
TREND PERIOD
Set to pre-power down setting
LOW SpO
2
ALARM LIMIT
Set to pre-power down setting
s e t u p
3-4
Radical Signal Extraction Pulse Oximeter Operator’s Manual