Masimo Radical Signal Extraction Operator'S Manual Download Page 29

Radical Signal Extraction Pulse Oximeter Operator’s Manual

4-23

4

o p e r a t i o n

NAVIGATING THE TREND  DISPLAY

In the Trend Display view there are a total of 10 touch key icon selections on 3 pages of
menu selections. These menu screens are not accessable when using the Simplified
User Interface.

FIRST PAGE

SECOND PAGE

THIRD PAGE

NEXT MENU
PAGE

Press the Next Menu Page button to return to the first page of
menu selections.

TREND SETUP

Press the Trend Setup button to enter the Trend Setup Menu.

HISTOGRAM

Press the Histogram button to display the selected data set (the
data set shown in the trend view) in histogram format.

CLEAR TREND
DATA

Press the clear Trend Data button to clear the data stored in the
trend memory.

NEXT MENU
PAGE

Press the Next Menu Page button to access the next page of
menu selections.

ZOOM

Press the Zoom button to change the time scale of the trend view.
The available time scales are 24 hrs, 12 hrs, 8 hrs, 4 hrs, 2 hrs, 1
hr, 30 minutes, 10 minutes, 1 minute and 20 seconds. The Zoom
button uses the last recorded data point as the zoom reference
point. In other words, the last recorded data point is always shown
as the right-most data point on the display.

ZOOM FROM
LEFT

Press the Zoom from Left button to zoom into the data set while
keeping the data point that is shown on the right side of the trend
graph as the zoom reference point.

ZOOM FROM
RIGHT

Press the Zoom from Right button to zoom into the data set while
keeping the data point that is shown on the left side of the trend
graph as the zoom reference point.

NEXT MENU
PAGE

Press the Next Menu Page button to access the next page of menu
selections.

EXIT

Press the Exit button to return to the normal display screen.

SCROLL
RIGHT

Press the Scroll Right button to scroll through the data set. The
display scrolls by ½ the selected time scale. For example if a 2 hr
display view is selected, then pressing the Scroll Right button will
scroll the displayed data by 1 hr to the right.

SCROLL
LEFT

Press the Scroll Left button to scroll through the data set. The
display scrolls by ½ the selected time scale. For example if a 2 hr
display view is selected, then pressing the Scroll Left button will
scroll the displayed data by 1 hr to the left.

By default, the trend display automatically refreshes, at a rate of once every 10 seconds, to
show the latest measured SpO

2

, PI or pulse rate data. This feature is only available while

the trend view is 2 hours or less, and the latest measured data is shown. If the user scrolls
through the data set to display previously recorded trend data, or if the trend scale is greater
than 2 hours, the trend display will time out after 1 minute of inactivity (i.e. the user does not
press any of the touch key control buttons) and the normal Radical display will be shown.

1

The top line on the trend display shows the time scale of the trend graph, followed
by the starting date, starting time and end time of the data set that is displayed on
the screen.

2

The second line of the display shows the minimum, average, and maximum SpO

2

,

PI or pulse rate measurements contained in the displayed data set (excluding zero
measurements).

3

The vertical scale of the SpO

2, 

PI or pulse rate graphs can be set in the Trend

Setup menu.

4

A dark line on the trend graph indicates the averaged data, while grayed-out data
points show minimum and maximum values.

5

A grayed-out box or line located on the bottom axis of the saturation graph indicates
a period of time for which the Low Signal IQ indicator was active, indicating the
signal quality was very low and the accuracy of the measurement may have been
compromised.

6

The available trend graphs show two of the desired trend parameter measurements
displayed versus time.

7

Masimo SET  

100

85

%SpO

2

140

50

BPM

10 Min         Trend                03/27/01   

11:42:46

% SpO

2

   90       97        99          BPM        60     70     80 

91

100

60

160

30

o p e r a t i o n

4-22

Radical Signal Extraction Pulse Oximeter Operator’s Manual

4

Summary of Contents for Radical Signal Extraction

Page 1: ...signal extraction pulse oximeter O P E R A T O R S M A N U A L...

Page 2: ...S i g n a l E x t r a c t i o n P u l s e O x i m e t e r O P E R A T O R S M A N U A L...

Page 3: ...00 Fax 949 297 7001 www masimo com EU Authorized Representative for Masimo Corporation MDSS GmbH Schiffgraben 41 30175 Hannover Germany Tel 49 511 62 62 86 30 Fax 49 511 62 62 86 33 MEDICAL ELECTRICAL...

Page 4: ...For home use ensure that the pulse oximeter s alarm can be heard from other rooms in the house especially when noisy appliances such as vacuum cleaners dishwashers clothes dryers televisions or radio...

Page 5: ...EW About this Manual 1 1 Product description 1 3 Features and benefits 1 3 Indications for use 1 3 Pulse oximetry 1 4 General Description 1 4 Principle of Operation 1 4 Functional saturation 1 5 Measu...

Page 6: ...o set up and use the Masimo SET Radical Pulse Oximeter herein known as the Radical Important safety information relating to general use of the Radical appears before this introduction Other important...

Page 7: ...roughout the hospital SpO2 pulse rate alarm trending perfusion index Signal IQ and plethysmographic waveform display Automatic screen rotation provides upright display for vertical or horizontal monit...

Page 8: ...te Saturation Transform DST reliably identifies the noise isolates it and using adaptive filters cancels it It then reports the true arterial oxygen saturation for display on the monitor P u l s e o x...

Page 9: ...tor instantly upgrading the conventional pulse oximetry to Masimo SET pulse oximetry The SatShare cable attaches to the back of the Radical Docking Station and SatShare cables are available to interfa...

Page 10: ...NSITIVITY The Max or APOD sensitivity icon is shown on the Radical display to indicate if the Radical is set to operate in Normal sensitivity Maximum sensitivity or Adaptive Probe Off Detection mode W...

Page 11: ...erfaces with a monitor through the SatShare cable The Output Mode Indicator will take place of the Time and Date indicator when utilized Philips Vuelink 20 POWER STAND BY BUTTON Press the Power Standb...

Page 12: ...dicator is illuminated when the Handheld unit is turned on and is properly interfaced to a Docking Station Check Sensor Masimo SET FastSat 11 42 46 03 27 03 100 85 SpO2 140 50 BPM 1 2 3 4 5 R a d i c...

Page 13: ...nector for AC power and two fuses The AC input provides power to the system from the AC line Always connect the pulse oximeter to the mains power for continuous operation and or battery recharging NOT...

Page 14: ...r cable to the power entry module of the Radical unit IEC 320 connector type at the Radical Connect the power cable to an AC power source Ensure that the unit is adequately powered by verifying that t...

Page 15: ...4 Set to off SENSITIVITY Set to normal mode IFTHERE IS ANY DOUBT ABOUTTHE INTEGRITY OFTHE PROTECTIVE EARTH CONDUCTOR ARRANGEMENT OPERATE THE OXIMETER ON INTERNAL BATTERY POWER UNTIL THE AC POWER SUPP...

Page 16: ...s and the Interface Alarms parameters in the General menu according to Customer preference 10 If displaying the simulated waveform is not desirable it is recommended to turn off the pleth waveform dis...

Page 17: ...o indicate the alarm conditions measured by the pulse oximeter the user must configure the RadNet Central Station monitor Please see the RadNet Hardwired and Wireless Systems Operator s Manual for com...

Page 18: ...Remove any substances that may interfere with the transmission of light between the sensor s light source and photodetector 5 Attach the sensor to the patient Refer to the Directions for Use of the se...

Page 19: ...NECTORS ARE NOT WATERPROOF DO NOT STERILIZE BY IRRADIATION STEAM AUTOCLAVE OR ETHYLENE OXIDE DO NOT ATTEMPT TO REPROCESS RECONDITION OR RECYCLE MASIMO SENSORS OR PATIENT CABLES AS THESE PROCESSES MAY...

Page 20: ...gnal quality of the measurement With motion the plethysmographic waveform is often distorted and may be obscured by artifact The Signal IQ shown as a vertical line coincides with the peak of an arteri...

Page 21: ...l displays a Low Perfusion message when there are very low amplitude arterial pulsations It has been suggested that at extremely low perfusion levels pulse oximeters can measure peripheral saturation...

Page 22: ...n 90 degree increments T o u c h k e y c o n t r o l b u t t o n a n d i c o n s The touch key control buttons are the four dark grey control buttons to the right of the Handheld display To select a t...

Page 23: ...ain menu is accessed the monitor remains functional and the saturation and pulse rate numbers will continue to be displayed MAIN MENU SELECTION The top menu category uses the following four menu selec...

Page 24: ...ing NOTE The SpO2 low limit can not be set below the password protected minimum low SpO2 alarm limit See Section 4 Operation Display for details PULSE RATE HIGH LIMIT BPM The pulse rate high alarm lim...

Page 25: ...Operator s Manual 4 MENU ITEMS DESCRIPTION VOLUME This menu allows the user to set the alarm volume Four levels are available level 1 being the softest and level 4 being the loudest The device retains...

Page 26: ...ion battery Selecting Yes will disable Docking Station functions such as SatShare Serial and Analog output Selecting No will activate these Docking Station functions while operating on battery power W...

Page 27: ...mally Open and Normally closed contacts See section 7 Analog output nurse call specifications O u t p u t Note The output menu selections are only available when the Radical Handheld is interfaced to...

Page 28: ...NIMUM OF 87 HOURS 3 5 DAYS 10 SECONDS TYPICALLY 435 HOURS 18 DAYS S e r v i c e NOTE The Service menu selections are only available when the Radical Handheld is interfaced to the Docking Station Only...

Page 29: ...play view is selected then pressing the Scroll Right button will scroll the displayed data by 1 hr to the right SCROLL LEFT Press the Scroll Left button to scroll through the data set The display scro...

Page 30: ...adical to store one set of SpO2 PI and pulse rate measurements every 2 seconds resulting in a minimum trend capacity of 87 hours A setting of 10 for example sets the Radical to store one set of data p...

Page 31: ...atShare mode Radical operates as follows All visual alarms remain active All audible alarms may be disabled by software configuration of the Radical The SpO2 and pulse rate numbers may or may not be d...

Page 32: ...ate operator response is required Low priority alarms indicate that operator awareness is required ALARM DESCRIPTION HIGH PRIORITY Audible and visual alarms for low saturation SpO2 range 1 100 and pul...

Page 33: ...n the saturation measurement or pulse rate measurement exceeds the low or high saturation or pulse rate alarm limits The Radical allows a user to delay audible saturation alarms by 0 5 or 10 seconds b...

Page 34: ...of the fuses have blown Replace the fuses UNIT POWERS ON BUT THE GRAPHIC DISPLAY IS BLANK The viewing contrast is not correct Use the Backlight Contrast button to adjust the viewing angle If the condi...

Page 35: ...w Perfusion3 Adults Pediatrics 2 digits Neonate 3 digits Pulse Rate Accuracy pulse rate 25 240 bpm No Motion1 Adults Pediatrics Neonate 3 digits Motion2 Adults Pediatrics Neonate 5 digits Low Perfusio...

Page 36: ...n limit to 0 02 S e r i a l i n t e r f a c e s p e c i f i c a t i o n s The digital interface for serial communication is based on the standard RS 232 protocol The Radical by default always outputs...

Page 37: ...2 SERIAL INTERFACE SETUP To interface with the Radical serial port set the following communication parameters on the interfacing serial device The pin outs for the RS 232 connector are shown in the f...

Page 38: ...infrared heating lamps and direct sunlight can interfere with the performance of an SpO2 sensor To prevent interference from ambient light ensure that the sensor is properly applied and cover the sens...

Page 39: ...LNOP REUSABLE SENSORS LNOP sensors must be used in conjunction with PC cables Note The LNOP TF I and TC I sensors were not validated under motion conditions LNOP ADHESIVE SENSORS LNOP sensors must be...

Page 40: ...from the patient 2 Disconnect the sensor from the monitor 3 Wipe the entire sensor clean with a 70 isopropyl alcohol pad 4 Allow the sensor to air dry before returning it to operation REATTACHMENT OF...

Page 41: ...er use a soft cloth dampened with a mild soap and water Do not allow liquids to enter the interior of the instrument CAUTIONS DO NOT AUTOCLAVE PRESSURE STERILIZE OR GAS STERILIZE THIS OXIMETER DO NOT...

Page 42: ...ld includes a 1 5 Amp Hour Nickel Metal Hydride battery The Radical Docking Station may include the optional 6 5 Amp Hour Nickel Metal Hydride battery Before using the Radical as a Handheld or transpo...

Page 43: ...rs or during routine maintenance follow the procedure outlined in this section If the Radical fails any of the described tests discontinue its use and correct the problem before returning the unit bac...

Page 44: ...as measuring data and verify that the resistance is less than 35 ohms SpO2 Pulse Rate High 100 140 Low 90 50 Power On Self Test 1 Connect the monitor to AC power and verify that the AC Power Indicator...

Page 45: ...MAY ONLY PERFORM MAINTENANCE PROCEDURES SPECIFICALLY DESCRIBED IN THIS MANUAL REFER SERVICING TO QUALIFIED SERVICE PERSONNEL TRAINED IN THE REPAIR OF THIS EQUIPMENT Analog Output Test 1 Disconnect al...

Page 46: ...f its agencies shall be governed solely by the terms of this Agreement and shall be prohibited except to the extent expressly permitted by the terms of this agreement W a r r a n t y Masimo warrants t...

Page 47: ...325 SATSHARE CABLE DS01 1539 SATSHARE CABLE DS02 1528 SATSHARE CABLE DS03 1357 SATSHARE CABLE HP03 1321 SATSHARE CABLE MQ01 1789 SATSHARE CABLE MS01 1367 SATSHARE CABLE NK01 1368 SATSHARE CABLE NK02 1...

Page 48: ...on on both the sensors and monitors to ensure accurate pulse oximetry when needed most 2007 Masimo Corporation Masimo Radical Discrete Saturation Transform DST Satshare SET LNOP LNCS and LNOPv are fed...

Reviews: