Radical Signal Extraction Pulse Oximeter Operator’s Manual
9-1
9
s e r v i c e a n d m a i n t e n a n c e
I n t r o d u c t i o n
This section covers:
■
How to test the operation of the Radical and the SatShare interface
■
How to properly clean the Radical
■
How to recharge and replace the batteries
■
How to replace the fuses
■
How to obtain service
Under normal operation, no internal adjustment or recalibration is required. Safety tests
and internal adjustments should be done by qualified personnel only. Safety checks
should be performed at regular intervals or in accordance with local and governmental
regulations.
C l e a n i n g
To clean the display panel, use a cotton swab moistened with 70% isopropyl alcohol and
gently wipe the panel.
To clean the outer surface of the oximeter, use a soft cloth dampened with a mild soap
and water. Do not allow liquids to enter the interior of the instrument.
CAUTIONS:
■
DO NOT AUTOCLAVE, PRESSURE STERILIZE, OR GAS STERILIZE THIS
OXIMETER.
■
DO NOT SOAK OR IMMERSE THE MONITOR IN ANY LIQUID.
■
USE THE CLEANING SOLUTION SPARINGLY. EXCESSIVE SOLUTION CAN FLOW
INTO THE MONITOR AND CAUSE DAMAGE TO INTERNAL COMPONENTS.
■
DO NOT TOUCH, PRESS, OR RUB THE DISPLAY PANELS WITH ABRASIVE
CLEANING COMPOUNDS, INSTRUMENTS, BRUSHES, ROUGH-SURFACE
MATERIALS, OR BRING THEM INTO CONTACT WITH ANYTHING THAT COULD
SCRATCH THE PANEL.
■
DO NOT USE PETROLEUM-BASED OR ACETONE SOLUTIONS, OR OTHER
HARSH SOLVENTS, TO CLEAN THE OXIMETER. THESE SUBSTANCES ATTACK
THE DEVICE’S MATERIALS AND DEVICE FAILURE CAN RESULT.
Refer to Section 8,
Cleaning and Reuse of Masimo Sensors for cleaning instructions of
the sensor.
WARNING:
ELECTRICAL SHOCK AND FLAMMABILITY HAZARD - BEFORE
CLEANING THE OXIMETER, ALWAYS TURN IT OFF AND
DISCONNECT THE POWER CORD FROM THE AC POWER
SUPPLY.