O3® Regional Oximeter
Chapter 4: Operation
www.masimo.com
30
Masimo
Display and Alarm Settings
There are two (2) ways to configure all parameter display and alarm settings:
1.
Press the parameter icon in the O3
®
Module window of the specific sensor. This
will take you directly to the settings options for the parameter. Select the
appropriate setting and press Ok to confirm. Press the
Home
button to return to
the main screen.
Or
2.
Press the "Gear" icon on the lower right corner of the main window to access the
Main Menu. Then press the "O3" icon to access the O3
®
Module menu. In the O3
®
Module menu select the "POD" icon to access the sensor menu. Press the
appropriate O3
®
Sensor for which to configure specific parameter display and
alarm settings.
Below are the various setting options for each parameter.
Note:
All setting options will hold through power cycle, unless otherwise stated.
rSO
2
Settings
The following are the various setting options for rSO
2
:
About
An informational read-only screen appears with the following definition for rSO
2
:
rSO
2
, displayed as a percentage (0 to 99%), is the measure of regional tissue oxygenation (at
the deep tissue level) local to the sensor site.
Alarms
Option
Description
Factory
Default
Configuration
Options
Resolution
High Limit Upper threshold of rSO
2
level that
triggers an alarm.
Off
2% to 99%
Off
1%
Low Limit Lower threshold of rSO
2
level that
triggers an alarm.
40%
1% to 98%
1%
Silence
Duration
Duration of the temporary suspension
of audible alarm for rSO
2
when the
"Alarm Silence" icon is pressed on
Root.
2m
30s, 1m, 2m,
5m
N/A