Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
2-1
2
Maintenance
2.1
Introduction
2.1.1
This chapter covers how to test the operation of the Radical-
7 and SatShare interface (if applicable), how to properly
clean the Radical-7 pulse oximeter, how to recharge and
replace the batteries, how to replace the fuses, and how to
obtain service under normal operation. No internal
adjustment or recalibration is required. Service should be
done by qualified personnel only. Safety checks should be
performed at regular intervals or in accordance with local
and governmental regulations.
WARNING: ELECTRICAL SHOCK AND FLAMMABILITY HAZARD.
BEFORE CLEANING THE OXIMETER, ALWAYS TURN
THE UNIT OFF AND DISCONNECT THE POWER
CORD FROM THE BACK OF THE UNIT.
2.2
Cleaning
2.2.1
To clean the display panel use a cotton swab moistened with
70% isopropyl alcohol and gently wipe the panel.
2.2.2
To clean the outer surface of the oximeter, use a soft cloth
moistened with a 70% isopropyl alcohol. Do not allow liquids
to enter the interior of the instrument.
2.2.3
To decontaminate unit, refer to Section 7.2.2.
CAUTIONS:
•
Do not autoclave, pressure sterilize, or gas sterilize this oximeter.
•
Do not soak or immerse the monitor in any liquid.
•
Use the cleaning solution sparingly. Allow unit to air dry. Excessive
solution can flow into the monitor and cause damage to internal
components.
•
Do not touch, press, or rub the display panels with abrasive cleaning
compounds, instruments, brushes, rough-surface materials, or bring
them into contact with anything that could scratch the panel.
•
Do not use petroleum-based or acetone solutions, or other harsh
solvents, to clean the oximeter. These substances attack the
device’s materials and device failure can result.