Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
4-3
4.2.15 To program Radical-7 Docking Station, perform the following
steps:
4.2.16 Connect a Docking Station to AC outlet and turn the switch
to ON position (if applicable).
4.2.17 Connect a serial cable between computer and the Docking
Station.
4.2.18 Install a known good Radical-7 Handheld to the Docking
Station.
4.2.19 Turn off the Handheld, if it is on.
4.2.20 Start the FlashApp program. Click Browse and select the
proper Image file to be loaded. It will usually be listed under
a SDSSA subdirectory.
4.2.21 Select the proper Port (usually COM1), and Select Docking
Station as the Target.
4.2.22 Press the “Green” Program button. A ‘status message’ of
“waiting for docking station” will be displayed.
4.2.23 Turn on the Handheld.
4.2.24 The green Program button will turn off, and the red Stop
button will illuminate. Wait until the green button turns on
again indicating a successful transfer. .
4.2.25 If the message on the Status window is “Complete –
Success”, continue with next step. If the message is
“Complete – Fail”, go back to 4.2.19 and repeat. If this cycle
is repeated 3 times without success, troubleshooting is
required.
4.2.26 To check software version numbers, perform the following
steps:
4.2.27 Turn off the Handheld.
4.2.28 Turn on the Handheld.
4.2.29 Press the Menu key.
4.2.30 Press the Down key and select “About”.
4.2.31 Verify that the Handheld (Handheld, Masimo SET) and
Docking Station (D-Station) software version(s) are the
intended version. The Handheld must be docked to the
Docking Station in order to see the version of the Docking
Station.