Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
6-5
Symptom
Cause and Corrective Action
Steps
1. Any front panel button
affected
2. Buttons do not activate
feature
3. Buttons hard to press
2. Replace Portable System Board
•
Rightmost Touch Key
Difficult to Depress
1. Open unit. If Swivel Connector flex
cable is interfering with button,
replace Swivel Connector
2. Replace Keypad.
•
"No Sensor", “Defective
Sensor” or “Sensor Off”
Displayed
1. Replace sensor.
2. Replace patient cable.
3. Replace oximeter board.
•
Display Does Not Rotate
Automatically
1. Ensure Docking Station is
powered up
2. Ensure Power Save feature is not
enabled
3. Replace Docking Station System
Board
•
Display Does Not Rotate
Manually
1. Flex interference? Replace Swivel
Connector.
2. Replace Keypad
3. Replace Portable System Board
•
Docking Station Fails
When Portable is Docked
1. Ensure software of Handheld and
Docking Station is compatible
2. Replace Portable System Board
3. Replace Docking Station System
Board
4. Replace Interface Cable
•
Fan Does Not Function
1. Ensure AC power is on
1. Replace Fan
2. Replace Docking Station System
Board
•
Portable Will Not Power
Up When Undocked
1. If Clicking is heard when
On/Off button is depressed,
perform Step 3.
1. Check NiMH Battery charge
2. Replace Battery if charging is not
successful
3. Replace Portable System Board.
•
Portable will not Power Up
When Docked
1. Docking Station is
functioning
2. Portable powers up
undocked
1. Replace Interface Cable
2. Replace Docking Station System
Board
•
Portable Will Not Turn Off
1. Replace Portable System Board
2. Adjust or replace Keypad