Rad-57 Signal Extraction Pulse CO-Oximeter Operator’s Manual
9-3
9
Alarm Limit Test
1. With the monitor turned on, select the Menu Access key and enter the Alarm menu.
Change the High Saturation Alarm parameter to a value two points below the currently
selected value, and accept the change.
2. Verify that the newly set parameter is shown on the Saturation Alarm Limit Display, next to the
SpO
2
or pulse rate measurement display.
3. Return the High Saturation Alarm parameter to its original setting.
4. Repeat steps 1 to 3 with the Low Saturation Alarm parameter.
5. Repeat steps 1 to 3 with the High Pulse Rate Alarm parameter.
6. Repeat steps 1 to 3 with the Low Pulse Rate Alarm parameter.
7.
Perform the above steps for the SpCO and SpMet alarm limits.
8. Reset the alarm limits again to the original settings.
LED Brightness
1. With the monitor turned on, select menu level 3 (see Section 4,
Setup Menu Level 3 - LED
Brightness and Factory Defaults
) and use the Up and Down Arrow keys to cycle through all 4
brightness levels.
2. Exit the Menu system by pressing the Enter key or waiting for the normal time-out.
S e r v i c e a n d R e p a i r
Repair Policy
Masimo or an authorized Service Department must perform warranty repair and service. Do not use
malfunctioning equipment. Have the Instrument repaired.
Please clean contaminated/dirty equipment before returning, following the cleaning procedure
described in Section 9,
Cleaning
. Make sure it is fully dry before packing the equipment.
To return the Rad-57 Instrument for service, please follow the Return Procedure.
WARNING: DO NOT REMOVE THE COVER OF THE MONITOR EXCEPT
FOR BATTERY REPLACEMENT. AN OPERATOR MAY ONLY PERFORM
MAINTENANCE PROCEDURES SPECIFICALLY DESCRIBED IN THIS MANUAL.
REFER SERVICING TO QUALIFIED SERVICE PERSONNEL TRAINED IN THE
REPAIR OF THIS EQUIPMENT.
Return Procedure
Please clean contaminated/dirty equipment before returning and make sure it is fully dry before
packing the equipment. Call Masimo at 800-326-4890 and ask for Technical Support. Ask for an RMA
number. Package the equipment securely – in the original shipping container if possible – and enclose
or include the following information and items:
■
A letter describing in detail any difficulties experienced with the Pulse CO-Oximeter. Please
include the RMA number in the letter.
S e r v i c e a n d M a i n t e n a n c e