4-6
Rad-57 Signal Extraction Pulse CO-Oximeter Operator’s Manual
4
O p e r a t i o n
■
Externally applied coloring (such as nail polish)
■
Elevated levels of bilirubin
■
Severe anemia
■
Low arterial perfusion
■
Motion artifact
Numeric Display - Pulse Rate
The Pulse Rate displayed on the Rad-57 may differ slightly from the heart rate displayed on ECG
monitors due to differences in averaging times. There may also be a discrepancy between cardiac
electrical activity and peripheral arterial pulsation. Significant differences may indicate a problem with
the Signal IQ due to physiological changes in the patient or one of the devices or application of the
sensor or patient cable. The pulsations from intra-aortic balloon support can be additive to the pulse
rate displayed on the Pulse CO-Oximeter. Inaccurate measurements may be caused by
■
Intravascular dyes such as indocyanine green or methylene blue
■
Arrhythmias
■
Intra-aortic balloon support