Press the arrow keys to change the displayed value.
Press the GLP/ACCEPT key to accept the value or the CAL key to escape.
Press the arrow keys to select the next/previous parameter.
Press the SETUP key to exit SETUP menu at any time.
The following table lists the SETUP parameters, their valid values range and the factory
settings (default):
Item
Description
Valid values
Default
CELL
Cell Constant
0.500 to 1.700
1.000
tdS
TDS Factor
0.40 to 0.80
0.50
tc
Temperature Coefficient
0.00 to 6.00%/°C
1.90
rEF
Reference Temperature
20.0 or 25.0 °C
25.0
TIME
Time (hh:mm)
00:00 to 23:59
00:00
DATE
Date (MM.DD.YYYY)
01.01.2000-12.31.2099 01.01.2005
bEEP
Beep Status
ON/OFF
OFF
bAud
Baud Rate
600; 1200; 2400; 4800;9600
2400
In Id
Instrument ID
0000 to 9999
0000
tEMP
Temperature Unit
°C or °F
°C