51
T
-250.0
-200.0
1.5
1.7
-200.0
0.0
0.5
0.7
0.0
400.0
0.2
0.4
E
-250.0
-200.0
1.0
1.2
-200.0
-100.0
0.3
0.5
-100.0
1000.0
0.2
0.4
R
1
0.0
200.0
1.7
1.9
200.0
1767.0
1.0
1.2
S
1
0.0
200.0
1.7
1.9
200.0
1767.0
1.1
1.3
B
1
600.0
800.0
1.5
1.7
800.0
1000.0
1.2
1.4
1000.0
1820.0
1.0
1.2
C
0.0
1000.0
0.5
0.7
1000.0
2316.0
1.5
1.7
XK
-200.0
800.0
0.2
0.4
BP
0.0
800.0
1.9
2.1
800.0
2500.0
0.6
0.8
L
-200.0
900.0
0.2
0.4
U
-200.0
0.0
0.4
0.6
0.0
600.0
0.2
0.4
N
-200.0
-100.0
0.8
1.0
-100.0
1300.0
0.3
0.5
CJC error outside of 23 ± 5°C is 0.05°C/°C
1. For thermocouple measure mode on B, R and S; round the specification up or down accordingly as there
is no resolution past the decimal point.
Table 17: RTD Read and Source
RTD Type
Range (°C)
Accuracy (°C)
Minimum
Maximum
Measure 4W/Source
Measure 3W
PT385, 10 ohm
-200.0
-80.0
1.3
2.5
-80.0
0.0
1.3
2.5
0.0
100.0
1.4
2.7
100.0
300.0
1.5
3.0
300.0
400.0
1.6
3.1
400.0
630.0
1.8
3.3
630.0
800.0
1.9
3.6
PT385, 50 ohm
-200.0
-80.0
0.3
0.5
-80.0
0.0
0.4
0.7
0.0
100.0
0.4
0.7
100.0
300.0
0.4
0.7
www.
GlobalTestSupply
.com
Find Quality Products Online at: