7. At the top of the display it will now read, “SW OPENED AT” and give you the pressure that
the switch opened at.
8. Press the “NEXT” option to view when the switch closed, and the dead band.
Important NOTE:
The previous example uses a normally closed switch. The basic
procedure is still the same for a normally open switch, the display will just read “OPEN”
instead of “CLOSE”.
psi
SW OPENED AT
MENU
28.000
LIGHT
30.000
psi
psi
SW TEST OPEN
MENU
28.000
LIGHT
------